No.11S031-03
26
STC-CL500E/CLC500E
Users guide Rev. 1.3
4.4.2 Description of the camera control commands (
The under line settings
are the factory default settings)
Command No.
10H:
MOD1[7..0]
D7:
0: Auto,
1: Manual
D6:
Trigger polarity
0: Positive,
1: Negative
D5:
Trigger mode
0: Edge preset,
1: Pulse width
D4:
Binning mode
0: OFF (Normal),
1: ON (Binning)
D3:
Scanning mode
0: Full scanning,
1: Partial scanning
D2 to D0: Reset mode
000: Non-Reset,
001: V-Reset,
010 to 111: No function
(Prohibited setting. Do not set these values)
11H:
MOD2[7..0]
D[7..0]
D7 to D4: No function
Sets always as "0000"
D3:
Function mode
0: Trigger mode,
1: Continuous mode
D2 to D0: Partial scanning
000: Full scanning,
001: 1/2 partial scanning,
010: 1/4 partial scanning,
111: Varialbe partial scanning,
011 to 110: No function
(Prohibited setting. Do not set these values)
12H:
MOD3[7..0]
D7 to D6: Video out
00: 10bit,
01: 8bit
10: 12bit,
11: No function
(Prohibited setting. Do not set these values)
D5:
0: Camera Link (CC1),
1: Power/IO connector
(No. 2 pin)
D4 to D3:
00: Normal,
01: Reserved trigger
10 to 11: H reset
D2 to D0: No function
Sets always as "000"
Continuous / Trigger mode
the manual (set as 1).
Sets the camera function mode.
D[7..0]
[The camera function mode 2] Initial data: 0
No video output without the trigger signal input while the camera works with the trigger mode.
Function mode is enable when the "Continuous/Trigger mode selection (MOD1[7])" is
Trigger signal input connector
Sets the camera function mode.
Exposure start mode
Command descriptions
[The camera function mode 1] Initial data: 9
Sets the camera function mode.
D[7..0]
[The amera function mode 3] Initial data: 0
D7
D6
D5
D4
D3
D2
D1
D0
D7 D6
D5
D4 D3 D2
D1
D0
D7
D6
D5
D4
D3
D2
D1
D0