Sentec AG, Ringstrasse 39, CH-4106 Therwil, Switzerland, www.sentec.com
HB-010103-e Date of release: 01/2022
8
4
Country specific example (CEE 7/7)
Plug the SDM mains cable into the wall socket. Do
not use extension cables or multipliers.
Ensure that the mains connection can be easily
disconnected from the mains supply.
Note:
As an additional means of protection,
Sentec recommends to perform a socket and
wiring test prior to use (see Annex).
5
Circuit
Breakers
In case the circuit breakers on the back of the
Isolation Transformer are triggered, they can be
reset by pushing them back into their lock position.
Note:
The Isolation Transformer has an overheat
protection inside, which responds at approx. 120°C
and switches off the transformer. The overheat
protection is self-resetting and will only switch on
the transformer again when it has cooled down.
Simplified depiction of final SDM & Isolation Transformer assembly
Instructed personnel from the homecare provider shall electrically test the Isolation Transformer at
least every 6 months (chapter 3.1).
Refer to
the Instruction Manual for the SDMS (HB-005771) or Technical Manual for the SDMS (HB-005752)
for additional information, e.g. transport and storage conditions.
Troubleshooting is provided in the ‘Service Manual for the SDMS’ (HB-005615)
available on
Leave this Instruction Manual (HB-010103) at the patient’s home during monitoring
(also available on
.
Provide the Lay User Manual (HB-010069-Directions for Lay Users) to the patient (also available on