2.2.

 

Wiring 

The cell wiring should be far away from power lines (motors, transformers) 

and placed in separate pipes. Soldered connections have to be applied in 

the junction box, (preferably screwed connections). It is advised to place a 

bag of SILICA GEL to keep dry inside  the junction box. SENSY could 

provide PVC junction box with a PG9 packing-gland, which could receive 4 

or 6 parallel cells. REF:   Junction box  

   

JBOX-4R (4 inputs - 1 output) 

   

JBOX-6R (6 inputs - 1 output) 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2.3.

 

Parallel wiring 

The cells must be installed in parallel, with the stranded mass wire joined to it. The sense must be joined to the cell supply, before the points of 

parallel wiring and the stabilising resistances.  

 

2.4.

 

Calibration 

It must be done after the sensor has been turned on for a while (10-15 minutes) to obtain a uniform temperature of the installation. The cells do not 

usually need to be adjusted with each other. However, when greater precision is needed, it is sometimes necessary to stabilise the cells individually 

with the resistances in the junction box. Those resistances are of several ohms (±10) and are installed in the supply circuit. A parallel adjustable 

resistance is mounted with a fixed resistance. The most sensitive cell will have its input resistance increased and the least sensitive will have its 

lowest input resistance. You will see that it is preferable to work on both supply cables: schematic mounting is given for your information and allows 

a variation of 0 to 20 ohms in series on the input impedance (2x10 ohms). 

 

Note: A well known weight of more than 20% of the nominal load of the system can be expected. The calibration error is always much higher than 

the error made on the evaluation of the load. 

 

A) Display (ex. Dv680) 

B) Junction box

 

C) Cell

 

R) Adjusting resistance 

 

1) – out measure (green) 

2) – in supply (yellow) 

3) + in supply (brown) 

4) + out measure (green) 

5) Ref. (sense) + (pink) 

6) Ref. (sense) + (grey) 

 

Содержание 5950 Series

Страница 1: ...Calibration 4 2 5 Measurement errors 5 2 6 Insulation test 5 2 7 Output impedance 6 2 8 Input impedance 6 3 MOUNTING WITH EASY MOUNT 7 3 1 Mounting with 3 load cells 7 3 2 Mounting with more than 3 loadcells 7 4 example of mounting without easy mount 8 4 1 Mounting with a rubber foot F5950 8 4 2 Mounting with a solid steel base A5950 8 5 USE IN POTENTIALLY EXPLOSIVE ATMOSPHERE OPTION 9 5 1 Intrins...

Страница 2: ...ensor 1 3 Electrical weldings When arc welding must be done on the structure it would be advised to install stranded ground wire in order that the derived current could not pass through the cell damaging it It would also be advised to disconnect the cells of the measuring instrument 1 4 Lightning If there is a risk of lightning it would be advised to isolate the cell completely and derive the curr...

Страница 3: ...o as not to exert any effort on the sensor 2 CABLING 2 1 Cable The cells are delivered with a 4 wire screened cable The screen shielded cable cannot in any case be in contact with the ground e g in metallic junction boxes it is necessary to isolate the screen with a sheath thermal The screen can only be connected to standardized earth It is advised to install a thermo retractable sheath retracted ...

Страница 4: ...stallation The cells do not usually need to be adjusted with each other However when greater precision is needed it is sometimes necessary to stabilise the cells individually with the resistances in the junction box Those resistances are of several ohms 10 and are installed in the supply circuit A parallel adjustable resistance is mounted with a fixed resistance The most sensitive cell will have i...

Страница 5: ...heet provided in the appendix and filled in beforehand 2 6 Insulation test The measuring of the insulating resistance is done with a multimeter The standardized testing voltage is 10 V It is applied to a conductor It can be determined by disconnecting the measuring instrument and applying voltage between one of the conductors and the metallic mounting structure or individually cell by cell to loca...

Страница 6: ...ent a resistance variation of several ohms would instead be a consequence of a severe overvoltage problem 2 8 Input impedance Input signal IN brown IN yellow its resistance is usually of 700 Ω 5 Ω its impedance must be in accordance with the individual cell data sheet If a different resistance is read it means that there is a break off or a short circuit current It is at the input that one finds d...

Страница 7: ... he distribution of loads and freedom for the dilations of the part to be weighed 3 2 Mounting with more than 3 loadcells When more than 3 load cells are used each one must be placed at the same level to obtain a good equivalent load distribution Each load cells output signals and the empty part to be weighed must be identical case of a symmetrical part So to avoid cramping the displacements dilat...

Страница 8: ...bsorb the shocks and the slights irregularities of the state of the ground surface 4 2 Mounting with a solid steel base A5950 This combination gives an easy solution for the installation when the anti rotation and counter force parts are incorporated to the structure The A5950 base has the necessary solidity to support the head of the sensor and allows it to move freely in the case of great dilata...

Страница 9: ...l elements it is mandatory to control if the sensor s output tension is still compatible with the electronic device in use and the requested accuracy See certificate for the special conditions for safe use 6 PERIODIC INSPECTIONS 1 Check output for zero load annually Output signal Min acceptable Max acceptable mV V 4 wires 0 15 mV V 0 15 mV V 4 20mA 2 wires 3 mA 6 mA 4 20mA 3 wires 3 mA 6 mA 0 5V 3...

Страница 10: ...15 184 6 12 18 Load impedance e Ω NA 750 1 000 5k Nominal sig range 0 1 2 mV V 4 20 mA 4 20 mA 0 1 5 V 0 1 10 V 10 0 10 V Saturation 3 mV V 24 mA 24 mA 11 V 1 Max 60 C for EX I T4 T6 and C6 options 2 5 to 12VDC for EX I T2 GD EX I T4 GD and EX I T6 GD options 3 9 28VDC for EX I C6 options 4 15 to 27VDC with a 1000 Ω bridge 8 GUARANTEE The manufacturer s guarantee is applicable as far as mounting r...

Страница 11: ...07 1 4 5950 B 7 5 15 t 59 38 5 54 62 9 7 89 60 0 08 0 15 1 4 5950 C 20 t 59 38 5 54 62 9 7 89 60 0 2 1 4 Rev 18 04 2018 Accessories Other capacities and dimensions available on request Load direction Wiring Dimensions in mm STANDARD DIMENSIONS 5950 TECHNICAL DRAWINGS LOW PROFILE COMPRESSION LOAD CELLS on request ...

Страница 12: ...AB 0 3 15 t 20 74 120 15 t 0 76 A5950 C 20 t 25 79 100 0 95 Material stainless steel Other view Other capacities and dimensions available on request STANDARD DIMENSIONS A5950 Dimensions in mm TECHNICAL DRAWINGS LOADING PLATES FOR 5950 Rev 16 04 2018 ...

Страница 13: ...m Ref Item Capacities I595x A 0 3 20 t x Material I5950 stainless steel I5955 alloy steel Electrical ground strap ACCESS I595x SEM Other views Other capacities and dimensions available on request TECHNICAL DRAWINGS EASY MOUNT FOR 5950 Rev 01 06 2018 ...

Страница 14: ...sions available on request Electrical ground strap Ref Item Capacities M595x A 0 3 20 t Material M5950 stainless steel M5955 alloy steel STANDARD DIMENSIONS M5950 Dimensions in mm Rev 16 04 2018 TECHNICAL DRAWINGS MOUNTING PLATES FOR 5950 ...

Страница 15: ...isted here under 2014 30 EU Electro Magnetic Compatibility Directive 2011 65 EU Restriction of the use of certain hazardous substances in the electrical and electronic equipment RoHS 2014 35 EU Safety low voltage directive Conception and compliance of this equipment is made according to all of part of the following standards EN 61326 2006 If designed manufactured and tested safety ref D DP SIL3 RE...

Отзывы: