Sensorex S675 Скачать руководство пользователя страница 3

1/4 TURN AND 
PULL TO REMOVE

Cable Considerations

To install the cable in conduit, a threaded cable cap assembly 
should be used (S653).  Before the main rigid conduit a length 
of flexible conduit (at least 6"longer than the insertion rod  
should be used so that the insertion rod can completely clear 
the adapter body (large body with 1" NPT that threads into ball 
valve) for removal or the electrode.

Part number 

 

Minimum Length Flexible Conduit

S675 

   24"(61cm)

S676 

   30" 

(76cm)

S677 

   36" 

(91cm)

FIG. 9

FIG. 10

BALL 

VALVE  

CLOSED

STEEL WASHER

O-RING 
FACE SEAL

ELECTRODE

WHITE 

RETAINER NU

T

CAUTIONS:  

1) DO NOT INSERT INSERTION ROD ASSEM-
BLY INTO SOLUTION WITHOUT ELECTRODE 
ATTACHED!!!
2) MAKE SURE ELECTRODE IS ORIENTED AS 
SHOWN IN FIG 4.  OR WITH GLASS POINTED 
DOWNWARD WHEN INSTALLING OR RE-
MOVING SO THAT LIQUID CANNOT FLOW 
BACK INTO THE INTERNAL CONNECTOR!

Insertion/Wet-Tap pH and ORP Assembly (No ATC)

Product Instructions

                                                                                                              PRODUCT INSTRUCTION SHEET 

Page 3 of 3

Parts covered by this product data sheet include: 

S675, S676, S677

Form:  InstallInsert [Rev: 2012-05-16] 

©2011 Sensorex Corporation