![Sensor SMT-200 Скачать руководство пользователя страница 78](http://html1.mh-extra.com/html/sensor/smt-200/smt-200_operation-and-technical-manual_1243987078.webp)
Section 5 Geophone Testing
Io
SENSOR
68
SMT-200 Operations and Technical Manual 2.00/R1
Frequency, Damping and Sensitivity Failures
If all of these parameters fail together then check for external noise
(electrical or mechanical). Ensure the noise test is selected and check the
noise level. These tests are grouped together as they are derived from the
same test. Noise is also likely to result in increased distortion.
Failure of frequency, damping or sensitivity on a particular string is likely to
be due to an element failure. The test procedure is similar to that just
described except a test is run and processed each time.
Resistance Failures
Resistance is directly related to temperature. If there is a continuous failure
on resistance but sensitivity is ok then it is possible the temperature of the
string is incorrectly set. Check the Smartec sensor is enabled and the
SMT-200 Bendix connector (containing the sensor) is at a similar
temperature to the string under test.
Check also the geophone interval and lead-in is correctly set.
Distortion Failures
Check the noise level is not excessive. Troubleshooting procedure is the
same as for continuous impedance. Check all geophones are securely
planted and are planted upright. A common cause for distortion failures is
the coil impacting the case end plate if there is an excessive tilt on the
phones. The manufacturers specification sheet will state the maximum angle
of tilt that is allowable.
If the elements natural frequency is less than 12 Hz then it is advisable to
run the test at constant excursion. For these phones the test may fail if the
distortion tests are run at the natural frequency without compensation.
--Test results--
No. 1
Temp 20.0 °C
Pol 1
Leak 1401 Mohm
Res 382 ohm
Freq *10.59 Hz
Damp *0.270
Sens *27.1 V/m/s
Dist 0.04 %
Imp 2205 ohm
Frequency High
0 Back to Main
- Repeat Test
Start next Test
--Test results--
No. 1
Temp 20.0 °C
Pol 1
Leak 1401 Mohm
Res *398 ohm
Freq 9.91 Hz
Damp 0.259
Sens 28.7 V/m/s
Dist 0.04 %
Imp 2205 ohm
Resistance High
0 Back to Main
- Repeat Test
Start next Test
--Test results--
No. 1
Temp 20.0 °C
Pol 1
Leak 1401 Mohm
Res 382 ohm
Freq 9.91 Hz
Damp 0.259
Sens 28.7 V/m/s
Dist *0.22 %
Imp 2205 ohm
Distortion High
0 Back to Main
- Repeat Test
Start next Test
Содержание SMT-200
Страница 2: ...Io SENSOR SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 11: ...Io SENSOR Section 1 Introduction SMT 200 Operations and Technical Manual 2 00 R1 1 Section 1 Introduction ...
Страница 103: ...Io SENSOR Appendix B Test Overview SMT 200 Operations and Technical Manual 2 00 R1 93 Appendix B Test Overview ...
Страница 104: ...Appendix B Test Overview Io SENSOR 94 SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 113: ...Io SENSOR Appendix C Parts List SMT 200 Operations and Technical Manual 2 00 R1 103 Appendix C Parts List ...
Страница 114: ...Appendix C Parts List Io SENSOR 104 SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 126: ...Appendix D Schematics Layouts Io SENSOR 116 SMT 200 Operations and Technical Manual 2 00 R1 D 1 2 SMT 200 ADC ...
Страница 127: ...Io SENSOR Appendix D Schematics Layouts SMT 200 Operations and Technical Manual 2 00 R1 117 D 1 3 SMT 200 Filter ...
Страница 128: ...Appendix D Schematics Layouts Io SENSOR 118 SMT 200 Operations and Technical Manual 2 00 R1 D 1 4 SMT 200 Keyboard ...
Страница 129: ...Io SENSOR Appendix D Schematics Layouts SMT 200 Operations and Technical Manual 2 00 R1 119 D 1 5 SMT 200 DAC ...
Страница 131: ...Io SENSOR Appendix D Schematics Layouts SMT 200 Operations and Technical Manual 2 00 R1 121 D 1 7 SMT 200 Power ...
Страница 132: ...Appendix D Schematics Layouts Io SENSOR 122 SMT 200 Operations and Technical Manual 2 00 R1 D 1 8 SMT 200 Display ...
Страница 135: ...Io SENSOR Appendix D Schematics Layouts SMT 200 Operations and Technical Manual 2 00 R1 125 D 3 2 UC3906 B1 ...
Страница 136: ...Appendix D Schematics Layouts Io SENSOR 126 SMT 200 Operations and Technical Manual 2 00 R1 D 3 3 UC3906 B2 ...
Страница 139: ...Io SENSOR Appendix E Quick Reference SMT 200 Operations and Technical Manual 2 00 R1 129 Appendix E Quick Reference ...
Страница 140: ...Appendix E Quick Reference Io SENSOR 130 SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 144: ...Appendix E Quick Reference Io SENSOR 134 SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 145: ...Io SENSOR Appendix F Addendum SMT 200 Operations and Technical Manual 2 00 R1 135 Appendix F Addendum ...
Страница 146: ...Appendix F Addendum Io SENSOR 136 SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 147: ...Io SENSOR Appendix F Addendum SMT 200 Operations and Technical Manual 2 00 R1 137 F 1 Addendum ...
Страница 148: ...Appendix F Addendum Io SENSOR 138 SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 152: ...Notes Io SENSOR 142 SMT 200 Operations and Technical Manual 2 00 R1 Notes ...