VALVE CONFIGURATION 

Seven out of the eight valve inlet connections on the back of the 

SMART CAL 360

 

are fully configurable (valve 6 is 

reserved for fresh air).  By default, valves will be setup for the most common configurations: 

Valve 1:  2.5%V/V Methane / 100ppm CO combo gas 

Valve 2:  Line Gas (for 100% V/V Methane calibration) 

Valve 3:  25ppm H2S 

Valve 4:  0.1% V/V (1000ppm) Methane 

Valve 5:  1.1% V/V Propane 

Valve 6:  Clean air (not configurable) 

Valve 7:  Clean air 

Valve 8:  100ppm Ethane 

If any of your instruments require calibration gas that isn’t listed above, or if 
you  need  to  change  the  existing  configuration,  configure  the  valves  by 

following the instructions below. 

1.  From the main display, select the menu button on the bottom right. 

2.  Enter  the  supervisor  password  (contact  SENSIT  if  you  need  the 

password).  

“Supervisor Menu” should appear at the top of the screen. 

3. 

Locate and select “Valve Configuration” from the list. 

4.  For each gas you want to setup, select one of the valves from the top of 

the  list,  whether  you  will  be  using  a  demand  or  positive  pressure 

regulator, and the appropriate gas from the list. 

5.  After all the valves are configured, hit the save icon on the bottom right.  

Hit the arrow in the upper left to return to the main display. 

 

 

 

Содержание SMART-CAL 360

Страница 1: ...ON STATION INSTRUCTION MANUAL READ AND UNDERSTAND INSTRUCTIONS BEFORE USE 851 Transport Drive Valparaiso IN 46383 8432 Email info gasleaksensors com Web www gasleaksensors com Phone 888 473 6748 219 465 2700 Fax 219 465 2701 ...

Страница 2: ...e to the metal fitting on the back of the station and the fresh air hose s to the appropriate Air port connection s shown as shown on screen c Attach the fittings on the hoses to the appropriate gas port connections on the back of the station as shown on screen If the configuration shown does not match your needs it can be re configured in the Supervisor Menu refer to the instruction manual and co...

Страница 3: ...e by a factory certified technician WARNING Follow the manual instructions and testing methods WARNING Only use in areas that are known to be free of combustible gas hazards WARNING Only for use in areas that are clean and dry WARNING Turn all regulators and or gas valves off when the SMART CAL 360 is not in use WARNING If your calibration setup has line gas plumbed directly into the station inste...

Страница 4: ...fications 8 Installation and Setup 9 Valve Configuration 10 Operation and Use 11 Performing a Calibration 11 Performing a Bump Test 12 Viewing Logs 13 SD Card Features 13 Database Backup and Restore 14 Software Updates 14 Supervisor Menu 15 Maintenance and Care 16 Filter Replacement 16 Troubleshooting 17 Flow Block Check Issues 17 Communication Issues 17 Check Gas Supply Errors 18 Notes 19 EU Wast...

Страница 5: ... 00064 Instruction Manual 750 00086 Calibrator protective dust cover 360 00636 SD Card 360 00243 SCAL 360 Data Management Software 500 SC360 01 DESKTOP VERSION USB Cable 360 00637 NETWORK VERSION Ethernet Cable 360 00027 Wi Fi Antenna 360 00481 ACCESSORIES AND REPLACEMENT PARTS SCAL 360 License Renewal 500 SC360 02 Gas Input Fittings with filter 360 00520 ...

Страница 6: ...e data on screen and perform bump tests The calibrator can be used as a stand alone device or with SCAL 360 Data Management Software See software setup section for specific requirements and instructions The calibrator automatically records all calibration and bump test attempts whether pass or fail to ensure 100 compliance The on board memory can store up to 6 000 calibration records and 12 000 bu...

Страница 7: ...f stations over a company s network regardless of their physical locations SCAL N 360 operates within the network and does not require an internet connection All stations share one central SQL server database allowing instrument users to calibrate bump test their instruments at any location The administrator using SCAL N 360 Admin adds stations and locations downloads data and can assign superviso...

Страница 8: ... Storage Temperature 32 F to 122 F 0 C to 50 C Maximum Input Pressure 30 PSI 2 Bar Supported Instruments as of 27 Mar 20 Legacy IR Communication SENSIT HXG 3 HXG 3P SENSIT Gold Gold 100 Gold CGI Trak IT III CGI SENSIT CO HCN SENSIT Gold G2 TRAK IT IIIa SENSIT PMD IRED Data Storage Capacity 6 000 Calibration Records 12 000 Bump Test Records 120 Gas Cylinder Logs If using an on demand style regulato...

Страница 9: ...rocess on the station is complete the main display should be shown On this display will be the default valve configuration which covers most standard calibration kits If you need to change this configuration based on the calibration gas you will be using for your instruments first follow the steps listed in the next section for valve configuration then proceed with these instructions 5 Attach the ...

Страница 10: ...ion gas that isn t listed above or if you need to change the existing configuration configure the valves by following the instructions below 1 From the main display select the menu button on the bottom right 2 Enter the supervisor password contact SENSIT if you need the password Supervisor Menu should appear at the top of the screen 3 Locate and select Valve Configuration from the list 4 For each ...

Страница 11: ...cessary check the instruction manual for your specific instrument 3 Place the instrument into the proper cutout position of the cradle and attach the instrument hose to the inlet of the instrument For Trak It instruments a pig tail assembly is required 4 On the SMART CAL 360 press the CALIBRATION button Searching will be displayed while the station and instrument establish communication After it h...

Страница 12: ...let of the instrument For Trak It instruments a pig tail assembly is required 4 On the SMART CAL 360 press the BUMP button Searching will be displayed while the station and instrument establish communication After it has been established and the model type and serial number of the instrument are displayed no other input is required Calibration will proceed and give a result after completion Bump t...

Страница 13: ...ill be displayed which organizes records by first by instrument type After the appropriate instrument type is selected all records for that type are displayed organized by date and time Each record will also show the instrument serial number whether the record is for a calibration or bump test and what the result was When a record is selected a breakdown is shown for the result of each test perfor...

Страница 14: ...eature enabled which means that once per day the station will automatically create and save a backup This will be done up to 10 times before the oldest backup is erased Backups can also be created manually from the Backup Restore Database menu in the Supervisor menu Contact SENSIT Technologies if you need assistance with this feature SOFTWARE UPDATES SMART CAL 360 software can be updated by the us...

Страница 15: ...IGURATION Resets the valve configuration to the default setup CYLINDER TRACKING Enable or disable cylinder tracking feature This allows the user to enter details about each gas cylinder attached to the station such as expiration date and lot numbers This information can then be retrieved by SCAL 360 Data Management software to send alerts related to gas usage ERASE CAL BUMP LOGS Password protected...

Страница 16: ... the back of the station Over time is it possible these filters may become clogged with debris or oils from inside the gas cylinders which would result in Check Gas Supply errors If this happens or if the filter looks damaged or frayed the filter will need to be replaced To replace a filter 1 Disconnect the tubing from the fitting and remove the old filter using a 9 16 15mm wrench or socket The pl...

Страница 17: ...ooks good place a finger over the inlet of the instrument Within 5 seconds a flow blocked message should appear on screen If that message does not appear or if it takes a significant amount of time the sampling system of the instrument isn t sealed properly If the instrument shows the flow block message properly check the instrument hose by attaching it to the instrument as you normally would and ...

Страница 18: ...You should hear and or feel a release of pressure 3 Inspect the filter fitting on the valve that is giving the error and the fitting on the hose that attaches to it Make sure the connection isn t loose and neither fitting has a visible crack If no obvious issues are found temporarily replace the filter either with a new one or from a valve position that isn t being used a Disconnect the tubing fro...

Страница 19: ...___________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ ___________________________________________________________________...

Страница 20: ...oducts with the WEEE Symbol figure above to alert our customers that products bearing this label should not be disposed of in a landfill or with municipal or household waste in the EU WARRANTY Your SMART CAL 360 is warranted to be free from defects in materials and workmanship for a period of two years after purchase If within the warranty period your instrument should become inoperative from such...