CALIBRATION 

continued from page 30

When the reading is satisfactory, the display will flash DATA
SAVED indicating that calibration is complete for that sensor.
The date for CAL PAST DUE is automatically reset for that sen-
sor as well.

Scroll with the SAVE/ZERO BUTTON (C) if you need to cali-
brate another sensor. When finished, remove and shut off the
gas supply. Press & release the POWER/MUTE BUTTON (A)
to return to the working display.

HYDROGEN SULFIDE (H2S) CALIBRATION  (H2S 25 PPM)
From the working display press & hold the BH TEST/MENU
BUTTON (B) until the top line reads USER MENU. Press &
release the SAVE/ZERO BUTTON (C) once. The bottom line will
read CALIBRATION.

Press & release the BH TEST/MENU BUTTON (B) once. The
bottom line will read CO/100 PPM. Press & release the SAVE/
ZERO BUTTON (C) once. The bottom line will read H2S 25ppm.
Apply 25ppm H2S/AIR to the instrument and press  & release
the BH TEST/MENU BUTTON (B) to start H2S calibration.

When the reading is satisfactory, the display will flash DATA
SAVED indicating that calibration is complete for that sensor.
The date for  CAL PAST DUE  is automatically reset for that
sensor as well.

Scroll with the SAVE/ZERO BUTTON (C) if you need to cali-
brate another sensor. When finished, remove and shut off the
gas supply. Press & release the POWER/MUTE BUTTON (A)
to return to the working display.

Manually zero the instrument prior to beginning the calibration
process.

NOTEUsing calibration kits other than recommended by
SENSIT TECHNOLOGIES  may cause inaccurate readings.
Repairs are required if any sensor fails to calibrate. Con-
sult SENSIT TECHNOLOGIES for details.

NOTE: When calibrating, the numbers shown on the dis-
play represent the numbers seen by the microprocessor
and should not be confused with actual gas readings.
These readings will update every 5 seconds during cali-
bration.

The following instructions pertain to manual calibration of the
Sensit Gold CGI. If you are using the automatic Smart-Cal Cali-
bration System, the procedure is different. See the Smart-Cal
sections of this manual (page 35) or consult the Smart-Cal in-
struction manual for details.

CARBON MONOXIDE (CO) CALIBRATION (100PPM CO/AIR)
From the working display press & hold the BH TEST/MENU
BUTTON (B) until the top line reads USER MENU.

Press & release the SAVE/ZERO BUTTON (C) once. The bot-
tom line will read CALIBRATION.

Press & release the BH TEST/MENU BUTTON (B) once. The
bottom line will read CO 100ppm. Apply 100ppm CO/Air calibra-
tion gas and press & release the BH TEST/MENU BUTTON (B)
to start CO calibration.

CALIBRATION

Содержание GOLD CGI

Страница 1: ...ered Units out of warranty will be repaired for a service charge Internal repair or maintenance must be completed by a SENSIT Tech nologies authorized technician Violation will void warranty Units must be returned postpaid insured and to the attention of the Service Dept for warranty or repair This warranty gives you specific legal rights and you may have other rights which vary from state to stat...

Страница 2: ... a potentially hazardous situation which if not avoided may result in minor or moderate injury NOTICE LEL sensor should be checked for accuracy after expo sure to any gases containing silicones high sulfur content high concentrations of propane and high concentrations of CO above 1000ppm or exhaust gases Continuously low calibration check results or fluctuation of zero read ings may indicate senso...

Страница 3: ...indicated by a red flashing LED display indicator and alarm GENERAL DESCRIPTION SENSIT GOLD CGI INSTRUMENTS MODEL CGI EX CGI EX CO CGI EX CGI EX CO CGI EX TOX CGI 4GAS Plus Plus Plus MODEL MODEL MODEL MODEL CGI EX TOX MODEL MODEL LEL ALARMS LEL GAS DISPLAY LEAK DETECTION CARBON MONOXIDE OXYGEN HYDROGEN SULFIDE SENSING FEATURES GENERAL DESCRIPTION continued from page 2 sound The combustible gas ala...

Страница 4: ...ck can be easily heard with the speaker located in the back of the instrument An infrared communication window is located on the right side to allow the Sensit Gold CGI instrument to download cali bration data readings the operator has elected to save to the on board memory as well as communicate with the Smart Cal Cali bration Station and PCI 2s computer interface A flexible gooseneck is used to ...

Страница 5: ...n altered sensor cap can cause damage to the SENSOR TYPES AND PUMPS 7 BATTERYINSTALLATION REPLACEMENT instrument and void the warranty Battery replacement is necessary when the display reads BAT LOW an audible alarm sounds and the green ready LED flashes When the instrument remains in BAT LOW a count down will appear starting at 300 seconds 5 minutes which is the maxi mum time remaining before shu...

Страница 6: ...n is 0 1 LEL the equiva lent of 50ppm methane As an option the instrument is capable of displaying a range of 0 1 LEL up to increments as high as every 2 LEL An alter native option is to request a PPM display Resolution for this option is 1ppm or every 10ppm An X on the display indicates that the instrument is not outfitted with that sensor 6 Prior to use test the integrity of the sensor cap and t...

Страница 7: ...THANE 50 LEL Methane to 17 volume Methane LED indicator only above 17 volume Methane PROPANE 50 LEL Propane to 12 volume Propane LED indicator only above 12 volume Propane b Oxygen below 19 5 and above 22 5 c Carbon Monoxide 35ppm per utility industry standards d Hydrogen Sulfide 10ppm and above per Federal OSHA guidelines Caution There are gases that can poison or be cross sensitive to the combus...

Страница 8: ...es During sampling the respective read ings may change Audible and visual alarms will activate when the preset limits are reached 15 If the instrument encounters a gas it is not calibrated to it may read NSR followed by a number If the instrument is calibrated for natural gas NSR Non Standard Reading likely indicates a heavy gas i e heavier than air such as gasoline propane carbon dioxide etc If t...

Страница 9: ...TON B to enter the BH menu Insert the bar hole probe into the location for the survey Press and release BH TEST MENU BUTTON B once more to start the test A 45 second countdown for the test will begin The current per cent of gas by volume detected will be displayed on the left BAR HOLE TEST BAR HOLE TEST continued from page14 The peak percent of gas by volume detected will be displayed on the right...

Страница 10: ...tain mod els ordered with the Extended Memory option AUTOLOG Automatic storage of peak gas readings of up to 1 600 events SHOW CO TEST Display CO levels recorded during timed test SHOW CF TEST Display calculated AIR FREE CO levels re corded during timed test PRINT MENU From the working display access the menu by pressing hold ing the BH TEST MENU BUTTON B until the top line of the display reads US...

Страница 11: ... the SAVE ZERO BUTTON C until the bottom line dis plays SET CLOCK Press release the BH TEST MENU BUT TON B once to enter the menu The day will be the section flashing on the display To change this section press release the BH TEST MENU BUTTON B for adjustments Press release the SAVE ZERO BUTTON C to advance to the next section month year or time Press release the POWER MUTE BUTTON A to save the se...

Страница 12: ... the SAVE ZERO BUTTON C to scroll to view the individual gas readings saved Press release the POWER MUTE BUTTON A once to access a different session number Press release the SAVE ZERO BUTTON C to scroll to a new session number The standard number of available stored sessions is factory set at 6 but is factory adjustable up to 100 To exit this menu press releasethePOWER MUTEBUTTON A toreturntothewo...

Страница 13: ...n If the sensor shows proper depletion within this pe riod PASSED will flash on the display Press release the POWER MUTE BUTTON A to return to the working display If the O2 sensor does not respond properly within the 20 second test FAILED will appear on the display Remove the instrument from service Consult the factory in the event of any failure Press releasethePOWER MUTEBUTTON A toreturntothewor...

Страница 14: ...y press release BHTEST MENU BUT TON B once SELECT TEST will appear on the top line of the display Press release the BH TEST MENU BUTTON B againto enter the CO test menu Press release the BH TEST MENU BUTTON B once more to start the test NOTE Using the SAVE ZERO BUTTON C may advance you to another test option depending on the instrument version A 180 second timed test will begin The time for this t...

Страница 15: ...s sampling of gas fired appliances MENU continued from page 26 NOTE The hot air flue probe must be used with the instru ment when conducting this test to prevent damage to the instrument and to receive proper calculations From the working display press release BHTEST MENU BUT TON B once SELECT TEST will appear on the top line of the display Press release the BH TEST MENU BUTTON B again and the ins...

Страница 16: ...and time of that test Press release the C again to view the readings of that test Invalid test data will show as N A for the peak CF level Data from previous test can be viewed by scrolling with the SAVE ZERO BUTTON C Press release the POWER MUTE 28 29 BUTTON A to return to the work display To Print a CF Test From the working display press hold the BH TEST MENU BUTTON B until the top line reads US...

Страница 17: ...h the SAVE ZERO BUTTON C if you need to cali brate another sensor When finished remove and shut off the gas supply Press release the POWER MUTE BUTTON A to return to the working display Manually zero the instrument prior to beginning the calibration process NOTE Using calibration kits other than recommended by SENSIT TECHNOLOGIES may cause inaccurate readings Repairs are required if any sensor fai...

Страница 18: ... for that sensor The date for CAL PAST DUE is automatically reset for that sen sor as well Scroll with the SAVE ZERO BUTTON C if you Continued on page 34 COMBUSTIBLE GAS CALIBRATION 50 LEL METHANE From the working display press hold the BH TEST MENU BUTTON B until the top line reads USER MENU Press release the SAVE ZERO BUTTON C once The bottom line will read CALIBRATION Press release the BH TEST ...

Страница 19: ...s The display will read SMART CAL communicating Please see the Smart Cal Calibration Station instruction manual for additional information need to calibrate another sensor When finished remove and shut off the gas supply Press release the POWER MUTE BUTTON A to return to the working display OXYGEN SENSOR TEST To determine if the O2 sensor is working properly verify the sensors reaction by exposing...

Страница 20: ...ff Alarm O2 low only Alarm CO Alarm H2S Cal Due Interval Show Session Log Show BH Test EL PPM 1 15 Up To 1600 0 30 Seconds 0 100 1 99 2 5 100 17 5 22 5 5 300ppm 2 30ppm 30 45 60 90 180 360 Days 1 15 1 15 0 100 by 1 up to 2 increments 0 2000ppm by 1ppm or 10ppm 6 Off 0 Seconds Off Off 50 17 CH 12 PRO 19 5 OSHA 35ppm 10ppm 30 Days 6 6 LEL by 0 1 4 Can be disabled ...

Отзывы: