AIRCON-2 HIGH VOLUME AIR SAMPLER
Sensidyne Document No. F-PRO-3100 (Rev D)
Table 3.8
Run Cycles
* NOTE (Step 2)
If you do not select a minimum of 1 cycle and enter ACCEPT, the unit will display a PROGRAMMING
ERROR message and return you to SELECT PROGRAM NUMBER. You will have to forward through the
time menus to # RUN CYCLES to re-enter your selection.
** NOTE (Step 3)
The collected RUN TIME (total time) equals the number of cycles times the programmed run time. The total
time cannot exceed 199:59 (hours:minutes). When you press ACCEPT, the unit automatically checks the
collected run time to see if it exceeds this limit. If it does, a PROGRAMMING ERROR message appears on
the display for 5 seconds. If this occurs, you must change the program RUN TIME and/or the # RUN
CYCLES be within the total time limit.
PROGRAMMING
#CYCLES
PROGRAMMING
#CYCLES
HOLD
SINGLE MODE
RUN TIME
1
2*
3**
4
PROG
ACCEPT
Begin set-up for RUN CYCLES.
Run Cycles position is flashing.
must select a minimum of 1 cycle
for the sampler to run.
Press and hold until the number
of cycles (00-99) is reached. You
Press to lock in RUN CYCLES.
Programming complete for
Program No. 1.
Puts unit in SINGLE MODE/
HOLD.
Step
Switch/Button
Action
Display Example