888
MGT Docking Station version II
3.2 Bump Test
Bump test is the process which checks and verifies the performance of the
monitors and ensures that sensors are correctly responding to the target gas.
To perform a Bump Test, follow the below:
✓
Insert the monitor(s) into the bay facing down.
✓
Slightly press down on the monitor until it locks into place.
✓
Press the Bump Test button for 2 seconds to execute Bump test.
✓
During the 20~30seconds, bay LEDs begin to light as the test progress.
✓
When test is finished, all (six) blue Bay LEDs illumination indicates if the MGT
has passed the test (see Figure 3).
✓
Otherwise, the third blue Bay LED illumination indicates if the MGT has failed
the test (see Figure 3).
Figure 3
MGT has passed
the test (six LEDs)
MGT has failed the
test (only 3
rd
LED)
Bay locking
mechanism
Note:
Do not use the monitors that failed the calibration test.
To remove the MGT monitors from the bay, pull to the left side bay
locking mechanism (lower-left side of each bay).
4. Trouble Shooting
4.1
Bay Light does not turn on /test not performed
✓
Clean Monitor IR window.
✓
IR interference (bright lights/sunlight).
✓
Monitor in gas alarm state.
4.2
All tests fail
✓
Gas cylinder empty
–
check pressure gauge if it indicates at least 100psi/10bar.
✓
Gas Expired
–
check gas expiration date located on the cylinder.
✓
Wrong gas concentration(s)
–
gas cylinder and MGT concentrations not same.
✓
Wrong gas type
–
check gas type.