Operating Instructions S1-E / S1-A
7 Conditioning
65 67 460 D3643
.
201.01.02.02
37
1. Brush the instrument under running water (< 38 °C, < 100 °F, at
least drinking water quality) and good lighting (min. 500 lux) and
color rendering index (min. 80 Ra) until no more dirt can be seen, for
at least 10 seconds.
2. Flush the drive channels with spray.
3. Conduct thermal disinfection or unwrapped steam sterilization.
4. Maintain mechanical parts manually [ → 38].
5. Maintain the push button chuck manually.
6. Pack the instrument in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
7. Sterilize the adapter [ → 43].