3. Take a picture of the unique ID on
the Main Unit. It is located on the
back of the Main Unit.
NB: The QR code sticker (the monitoring unit’s unique
ID) is now on the Main Unit casing itself, and not on
the backplate as shown in the pictures.
The picture of the unique ID on the
Main Unit should be sent together
with the installation address to the
service company monitoring the
AEDs after the installation has been
completed. Alternative or additional
steps may be required depending on
the service company.
4. Open the cabinet and place the
cable-tie mount on the right side of
the cabinet, centered and about 12
cm from the top.
NB: When the cabinet is opened
you’ll hear a high-pitched alarm for
about 30 seconds. You can stop the
alarm by using a magnet placed on
the door contact. You need to bring
your own magnet for that purpose.
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020)
7 of 15