INDICATOR PARAMETER SETTINGS
The parameter settings menu has a calibration section (C01 to C07 explained above) and a
parameter settings section (C08 and up).
To access the calibration section the seal switch (located at one corner of the PCB) must be
OFF. This will allow access to all C01 and up settings. If the seal switch is ON, then only C08 and up
can be accessed by the user. If you break the official seal by opening the back of the indicator to
access the seal switch, you may need to have the indicator recertified. Be sure to adjust the seal
switch back to the original setting after calibration/configuration has been performed.
To enter calibration/parameter settings, follow the procedure below:
1.
Press and hold the SET and ESC key at the same time to enter the parameter settings
2.
Navigate through the settings (C01 to C45) by using the number keypad
3.
Press the
#
key to enter/edit the parameter setting
Press the
ESC
key to save and exit settings at any time
Table 1. Indicator Parameter Settings
Function
Parameter Settings/Options
Weighing Unit
C01
1 = kg
2 = lb
Decimal Setting
C02
0 = no decimal
1 = #.#
2 = #.##
3 = #.###
4 = #.####
Graduation Setting
(readability of the least
significant digit)
C03
options: 1/2/4/10/20/50
Example with no decimal places (ie C02=0)
1 = 1 lb
2 = 2 lb
5 = 5 lb
10 = 10 lb
20 = 20 lb
50 = 50 lb
Maximum Capacity
C04
set max capacity ex. 100kg = 0100.00
Zero Calibration
C05
0 = zero calibration not needed
1 = set the zero calibration (Please ensure scale is empty and the
stable light is on)
Calibration
C06
0 = calibration not needed
1 = ready to calibrate with calibration weight
Restore Default
Settings
C07
0 = do not restore
1 = restore to default settings