11
CALIBRATION cont.
23. The C6 channel is used to calibrate the scale with a known weight. Press ZERO to
set the value of C6 to [
C6 1
]. Press SET
. The display will flash
SPAN
, and
then show [######].
24. Enter the calibration weight value you will use (at least 10% of max capacity you set
in C04 by using HOLD and TOTAL to move the cursor left and right, and ZERO
and
UNITS and to move the values up and down.
25. Place the calibration weight you have on the empty scale and press SET
.
26. The scale will count down from 10 to 0. Once 0 has been reached, the display will
show
CALEnd
.
27. Press SET
to continue. The display will now show
C07
.
28. Press ON/OFF
to save and exit the setup menu.
29. The scale has now been calibrated. The display will show the value of the calibration
weight on the scale.
30. If the scale does not show the value of the calibration weight, check that the feet on
the platform are not screwed in too tightly, and verify that the platform is level.
31. Unload the scale; the display should read 000000.
32. If the scale does not display 00000, check that the feet on the platform are not
screwed in too tightly, and verify that the platform is level.