b.
Rectal use
: Lubricate silver probe tip with petroleum jelly for easy insertion.
Gently insert sensor approximately 1cm (less than 1/2 ") into rectum. Normal
temperature between 36.2 and 37.7
℃
(97.2
℉
and 99.9
℉
).
c.
Armpit Use
: Wipe armpit dry. Place probe in armpit and keep arm pressed
firmly at side. From a medical viewpoint, this method will always provide
inaccurate readings, and should not be used if precise measurements are
required. Normal temperature between 35.2
℃
and 36.7
℃
(95.4
℉
and
98.1
℉
).
3.
The degree sign flashes throughout the testing process. When flashing stops an
alarm will beep for approximately 10 seconds. The measured reading will appear on
the LCD simultaneously. The minimum measurement time until the signaling tone
(beep) must be maintained without exception. The measurement continues even
after the buzzer notification. So that in order to achieve better body temperature
measurement result, recommend to keep the probe in mouth and rectum about 2
minutes, or in armpit about 5 minutes regardless of the beep sound and at least 30
seconds measurement interval should be maintained.
*Note: Normally the buzzes are “Bi-Bi-Bi-Bi-”; Alarm beeps more rapidly when
temperature reaches 37.8 (100
℉
) or higher, and the buzzes are « Bi-Bi-Bi-------Bi-Bi-
Bi-------Bi-Bi-Bi »
4.
To prolong battery life, press the On/Off Button to turn unit off after testing is
complete. If no action is taken, the unit will automatically shut off after around10
minutes.
TROUBLESHOOTING
Error message
Problem
Solution
Temperature taken is
lower than
32.0°C(90.0°F)
Turn off, wait one minute and take a new
temperature via close contact and sufficient
rest
Temperature taken is
higher than
42.9°C(109.9°F)
The system is not
functioning properly
Unload the battery, wait for 1 minute and
repower it. If the message reappears,
contact the retailer for service
Dead battery: Battery
icon is flashing, can’t be
measurable
Replace the battery
BATTERY REPLACEMENT
1.
Replace battery when « » appears in the lower right corner of LCD display.
2.
Pull battery cover off as shown in Figure 3.
3.
Gently pull out plastic circuit board with battery chamber approximately 1 cm
(slightly less than1/2") (See figure 4).
4.
Use pointed object such as a pen to remove old battery. Discard battery lawfully.
Replace with new 1.5V DC button type LR41 or SR41, UCC392, or equivalent. Be sure
battery is installed with polarity facing up. (see figure 5)