background image

b.

 

Rectal use

: Lubricate silver probe tip with petroleum jelly for easy insertion.  

Gently insert sensor approximately 1cm (less than 1/2 ") into rectum.  Normal 

temperature between 36.2 and 37.7

 (97.2

 and 99.9

). 

c.

 

Armpit Use

: Wipe armpit dry.  Place probe in armpit and keep arm pressed 

firmly at side. From a medical viewpoint, this method will always provide 

inaccurate readings, and should not be used if precise measurements are 

required. Normal temperature between 35.2

 and 36.7

℃ 

(95.4

 and 

98.1

). 

3.

 

The degree sign flashes throughout the testing process.  When flashing stops an 
alarm will beep for approximately 10 seconds.  The measured reading will appear on 

the LCD simultaneously. The minimum measurement time until the signaling tone 

(beep) must be maintained without exception. The measurement continues even 

after the buzzer notification. So that in order to achieve better body temperature 

measurement result, recommend to keep the probe in mouth and rectum about 2 

minutes, or in armpit about 5 minutes regardless of the beep sound and at least 30 

seconds measurement interval should be maintained. 

*Note: Normally the buzzes are “Bi-Bi-Bi-Bi-”; Alarm beeps more rapidly when 

temperature reaches 37.8 (100

) or higher, and the buzzes are « Bi-Bi-Bi-------Bi-Bi-

Bi-------Bi-Bi-Bi » 

4.

 

To prolong battery life, press the On/Off Button to turn unit off after testing is 

complete.  If no action is taken, the unit will automatically shut off after around10 

minutes. 

 

TROUBLESHOOTING 

Error message 

Problem 

Solution 

 

Temperature taken is 

lower than 

32.0°C(90.0°F) 

Turn off, wait one minute and take a new 

temperature via close contact and sufficient 

rest 

 

Temperature taken is 

higher than 
42.9°C(109.9°F) 

 

The system is not 

functioning properly 

Unload the battery, wait for 1 minute and 

repower it. If the message reappears, 

contact the retailer for service 

 

Dead battery: Battery 

icon is flashing, can’t be 

measurable 

Replace the battery 

 

BATTERY REPLACEMENT 

1.

 

Replace battery when «   » appears in the lower right corner of LCD display. 

2.

 

Pull battery cover off as shown in Figure 3. 

3.

 

Gently pull out plastic circuit board with battery chamber approximately 1 cm 

(slightly less than1/2") (See figure 4). 

4.

 

Use pointed object such as a pen to remove old battery.  Discard battery lawfully.  

Replace with new 1.5V DC button type LR41 or SR41, UCC392, or equivalent.  Be sure 

battery is installed with polarity facing up. (see figure 5) 

5.

 

Slide battery chamber back into place and attach cover. 

Содержание MT-4333

Страница 1: ...wet as inaccurate readings may result Do not bite the thermometer Doing so may lead to breakage and or injury Do not attempt to disassemble or repair the thermometer Doing so may result in inaccurate readings After each use disinfect the thermometer especially in case the device is used by more than one person Do not force the thermometer into the rectum Stop insertion and abort the measurement wh...

Страница 2: ...ronment and complies with the requirements of EN 60601 1 2 EMC IEC EN60601 1 Safety standards And the manufacturer is ISO 13485 certified CONTENTS 1 Thermometer 1 Owner s Manual 1 Storage Case PRODUCT ILLUSTRATION PRECAUTION The performance of the device may be degraded should one or more of the following occur Operation outside the manufacturer s stated temperature and humidity range Storage outs...

Страница 3: ...1060hPa Storage and transportation condition Temperature 20 55 4 131 Humidity 15 95 RH Atmospheric Pressure 700hPa 1060hPa Ingress Protection Rating IP27 Classification Type BF SWITCHABLE Temperature readings are available in the Celsius or Fahrenheit scale located in the upper right corner of LCD With the unit off press and hold the On Off Button for approximately 2 seconds to change the current ...

Страница 4: ...buzzes are Bi Bi Bi Bi Alarm beeps more rapidly when temperature reaches 37 8 100 or higher and the buzzes are Bi Bi Bi Bi Bi Bi Bi Bi Bi 4 To prolong battery life press the On Off Button to turn unit off after testing is complete If no action is taken the unit will automatically shut off after around10 minutes TROUBLESHOOTING Error message Problem Solution Temperature taken is lower than 32 0 C 9...

Страница 5: ...hly Finish by wiping with a soft dry cloth For disinfection 75 Ethanol or Isopropyl alcohol can be used Observe the following to prevent damage to the thermometer Do not use benzene thinner gasoline or other strong solvents to clean the thermometer Do not attempt to disinfect the sensing section tip of the thermometer by immersing in alcohol or in hot water water over 50 122 Do not use ultrasonic ...

Страница 6: ...The product is in compliance with the requirements of MDD 93 42 EEC 0197 is the identification number of notify body 0197 ...

Отзывы: