66
K100000000S9 036103 130121
2
Once inserted the new ID code in the instrument,
unscrew the four fixing screws of the instrument
base; then slide the base of the instrument to access
the internal parts.
3
Locate the position of the sensor to be replaced; be-
low is an example of a sensor to be replaced.
1
Before proceeding with the replacement of the sensor, it is necessary to insert in the instrument the ID code of
the new sensor, after installing the appropriate PC software downloadable from the website
www.seitronamericas.com.
4
Remove the tubes from the sensor to be replaced;
below is an example of a sensor with the
disconnected pipes.
5
After disconnecting the pipes, pull the sensor
upwards; below is an example of the sensor
compartment without the sensor.
16.3 Gas sensors replacing
In order to be authorized to the sensors replacing, it is necessary to install on the PC the “Easy2Swap” program;
to obtain this program send an e-mail to this address: [email protected]. The procedure for the sensor
replacement is described in detail on the manual which comes with the Software.
Note: the metrological chain validation can be only obtained with the calibration through referable samples.