![SEFRAM MW 3031 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/sefram/mw-3031/mw-3031_user-manual_1233359025.webp)
25
sine wave at full scale and non-sine wave
below half scale. For non-sine wave add the
following Crest Factor corrections :
For Crest Factor of 1.4 to 2.0, add 1.0% to
accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to
accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to
accuracy.
Max.CF of 400.0V range :
2.5
Max.CF of 600V range :
1.7
Input Impedance :
10M
Ω, <
100pF.
Overload protection :
600V rms.
DCV
Range
Resolution
Accuracy
400.0V
0.1V
± (
0.6%+ 2 D)
600V
1V
± (
0.7%+ 5 D)
Input Impedance :
Approx. 10M
Ω, <
100pF.
Overload protection :
600V rms.
Resistance / Continuity
Range
Resolution
Accuracy
400.0
Ω
0.1
Ω
± (
0.9%+ 5 D)
4.000k
Ω
0.001k
Ω
± (
0.9%+ 2 D)
40.00k
Ω
0.01k
Ω
400.0k
Ω
0.1k
Ω
4.000M
Ω
0.001M
Ω
± (
1.5%+ 5 D)
40.00M
Ω
*
0.01M
Ω
* There is a little rolling less than
±
2%
Open Circuit Voltage :
Approx. 0.4V