Page 8
Quick reference guide USB-MESY
HID 570.77A
\ 10.2013 \
Take the USB connecting cable and connect the
USB-MESY HID with your PC.
Connect the thermocouple sensors of type K with the
measurement system.
Start the software “Mesy for Windows”.
Choose in the pull down menu Mesy/… …/Adjust-
ment".
With the button [Reset] you can reset the adjust-
ments values back to the standard settings at any
time. After confirmation the values will be reset.
You can also read out the already existing adjust-
ment values by pressing the button [Read].
Provide two pots filled with an insulating fluid with
two different temperatures, for example 25°C and
100°C.
4.2. Adjustment of the measurement box
Thermocouple sensors are available with different accuracies.
With this function you can adjust the USB-MESY HID for each
connected sensor.
This adjustment is a two point adjustment. Temperatures mea-
sured with the MESY will be compared with a second calibrated
temperature value.
Press the button [On/Off] until you hear two long
beeps to switch off the MESY.
Attention:
A starting of a new measurement will overwrite
the previous recorded values
.
Connect the USB-MESY HID with your PC to read
the recorded measurement values. The MESY will
switch on automatically.
Start the software “Mesy for Windows” and press the
button [2] START to read the recorded values. The
values will be readout from the memory of the USB-
MESY HID and will be shown on the monitor.