c.  Affix the alarm and holder securely to back of 

wheelchair out of sight and reach of resident. Simply 
wrap the Velcro

®

 strap with the rubberized non-slip 

backing attached to the back of the holder tightly 
around wheelchair frame. 

CAUTION: Do not affix 

alarm to a PVC rail. The strap may permanently 
stain the rail. 

Holder placement should be easily 

accessible to staff. Secure excess cord to avoid 
damaging cord. 

d.  See Under Seat Adjustable Switch Bar instruction 

sheet for bar set up.

e.  To test system press either the “Test/Reset” button 

on the front of the alarm or apply pressure on the 
seat of the wheelchair after installing the Under Seat 
Adjustable Switch Bar. To stop alarm from sounding 
simply press the “Test/Reset” button on the front of the 
alarm 3 times quickly in 1.5 seconds. 

f.  Carefully assist resident into wheelchair. If using a seat 

cushion there should not be any resident discomfort 
when seated. If the resident is sitting directly on the 
seat, the top of the switch bar my cause discomfort 
to resident’s buttocks depending on the resident’s 
weight. The discomfort can be eliminated by either a 
seat cushion or lowering the switch bar as directed 
in instruction sheet included with the Under Seat 
Adjustable Switch Bar. 

g.  The alarm will allow for normal movement. However, if 

resident attempts to exit the wheelchair the alarm will 
sound to alert staff. When resident sits back down the 
alarm will stop sounding and automatically reset. 

h.  However, the light bar will continue to flash red 

intermittently to let staff members know which resident 
activated the alarm. The staff member needs to firmly 
press the “Test/Reset” button 3 times quickly in 1.5 
seconds to stop flashing light. The alarm is then 
automatically reset.

i.  When staff removes resident from wheelchair for 

whatever reason the alarm will sound. To silence the 
alarm simply press the “Test/Reset” button on front of 
alarm. 

NOTE: As a tamper resistant feature please 

press the “Test/Reset” button 3 times quickly in 1.5 
seconds to stop alarm from sounding.

j.  When the resident is positioned back in the wheelchair 

the alarm will beep twice signaling that the alarm has 
automatically reset.

k.  The alarm and switch bar should be checked prior 

to each use for proper functioning. Please note: The 
system can be checked while resident is on the switch 
bar by simply pressing “Test/Reset” button. The alarm 
will beep to indicate proper alarm functioning. 

DO 

NOT ASSUME! TEST EACH TIME THE SYSTEM IS 
USED!

l.  The light bar on the front of the alarm also functions 

as a low battery indicator. When the batteries are 
low and require changing the light bar will flash blue 
intermittently and the alarm will emit a beeping sound. 

Note: As a safeguard simply pressing the reset 
button will not stop the alarm from beeping when 
the batteries are low. The blue light will continue to 
flash and the beeping sound will continue until the 
batteries are changed.

 

Please note: For resident safety, the Under Seat Alarm 
does not have an “On/Off” switch. Staff members do 
not have to remember to turn alarm back on as alarm 
has an auto reset feature.

12 MONTH LIMITED WARRANTY

Personal Safety Corporation warrants this product to be 
free from factory defects in materials and workmanship for 
a period of 12 months from the date of purchase. Void if 
this product has been misused, abused or punctured. No 
warranty of responsibility of use expressed or implied.

 Revised: 08/22/18

WARNING!

Must remove switch bar and alarm before cleaning wheelchair or using wheelchair when showering 

resident. Failure to due so will damage system and will void warranty. 

LOW BATTERY INDICATOR

Test / Reset

VOICE                  ALARM

Secure

®

 Hip Protectors - SHP 

(In five sizes)

Secure

®

 Reversible Safety  

Floor Mat - MAT-1 & MAT-2

Secure

®

 Voice Alarm  

SVA-1

Secure

®

 Universal Alarm  

SUA-1

Please see our complete line of Secure

®

 fall & wandering prevention products at www.PadAlarm.com

Please see our complete line of Secure

®

 fall/wandering prevention & patient safety products  

at www.SecureSafetySolutions.com 

Highest Quality...Lowest Prices...Guaranteed!

Personal Safety Corporation  •  Hiawatha, Iowa 52233  •  800.373.2873

Отзывы: