INW LEVELSCOUT INSTRUCTIONS
Seametrics • 253.872.0284
Page 17 inwusa.com
Appendix B: Settings and Calibration
Settings and calibration values can only be changed when there is no data on the
sensor. Be sure to retrieve any data and then erase the data on the sensor before
proceeding.
The temperature channel rarely needs adjustment. If you think your temperature
channel needs calibrating, contact your service representative.
Before leaving the factory, your LevelSCOUT has been inspected using precision
instruments. However, you may want to change some of the settings for the level
channel for the following reasons:
•
To set a specific zero-reference point for submergence
• To return values as depth-to-water
• To return values as elevation in relation to sea-level
• To return values relative to a staff gauge
Preparation
• Run Aqua4Plus Lite or Aqua4Plus and scan for sensors.
• On the Sensor Map, click the sensor you want to adjust.
• From Aqua4Plus Lite:
◦
Select the units you want to use for measurements, i.e., psi, feet of water, inches
of mercury, etc. Set this from the Configure Menu | Program Configuration | Set
Computer Display Units.
◦
Open the Settings and Calibration window by clicking on the
Settings button.
• From Aqua4Plus:
◦
Select the units you want to use for measurements, i.e., psi, feet of water, inches
of mercury, etc. Set this from the Options Menu | Display Units.
◦ Open the Settings and Calibration window from the Configure Menu |
Settings and Calibration.
Select the setting type from the Select Setting Type dropdown box. A picture will display
illustrating the type of selection. Step-by-step instructions will be displayed on the right.
SETTINGS AND CALIBRATION