Problem
Probable Cause
Try...
Laptop screen reads
”No connection” and “Ac
-
tive” light is NOT blinking
Laptop screen reads
”No connection,” but “Ac
-
tive” light IS blinking
Flow rate incorrect
No flow rate
Wait for “Active” light to blink
Replace battery
Tighten cable
Test cable (tech bulletin “DL75/76: Testing the
DC3 Serial Cable” on Downloads page at
www.Seametrics.com)
Make sure K-factor in DL76 matches flow sen
-
sor
Mechanical: See if rotor spins freely
EX Magmeter: See if “Status LED” indicator is
blinking
Check for flow rate
DL76 is transmitting data.
Wait two minutes before taking action.
Dead battery in DL76
Loose cable
Bad cable
K-factor setting is incorrect
Damaged flow sensor
Broken wire