3.5. TRANSISTOR 

TESTING

3.5.1.

Set the rotary switch to the ‘hFE’ position.

3.5.2.

Determine whether the transistor to be tested in NPN or PNP type and locate the Emitter, Base and Collector leads. Insert leads of the transistor into

the correct holes in the transistor testing socket.

3.5.3.

The meter will show the approximate hFE value at test conditions of base current 10µA and Vce 2.8V.

4. SPECIFICATION

DC VOLTAGE

Range

Accuracy          Resolution

200mV

±0.25% of reading ±2 digits

0.1mV

2V

±0.5% of reading ±2 digits

1mV  

20V

±0.5% of reading ±2 digits

10mV

200V

±0.5% of reading ±2 digits

0.1V

1000V

±0.5% of reading ±2 digits

1V

Overload protection: 220Vrms AC for 200mV range

and 1000V DC or 750V rms AC for other ranges.

AC VOLTAGE

Range

Accuracy             Resolution

200V

±1.2% of rdg ± 10 digits

100mV

750V

±1.2% of rdg ± 10 digits

1V

Overload protection: 1000V DC or 750V rms. as 

for all ranges.

Frequency range: 45Hz - 450Hz

Response: Average calibrated in rms sine wave.

RESISTANCE

Range

Accuracy                Resolution

200

±0.8% of reading ±2 digits

0.1

2k

±0.8% of reading ±2 digits

1

20k

±0.8% of reading ±2 digits

10

200k

±0.8% of reading ±2 digits

100

2M

±1.0% of reading ±2 digits

1k

Maximum open circuit voltage: 2.8V

Overload protection: 15 seconds at 220V DC or rms.

as for all ranges.

DC CURRENT

Range

Accuracy

Resolution

200µA

±1% of reading ± 2 digits

100nA

2mA

±1% of reading ±2 digit

1µA

20mA

±1% of reading ±2 digits

10µA

200mA

±1.2% of reading ±2 digits

0.1mA

10A

±2.0% of reading ±2 digits

10mA

Overload protection: F200mA/250V fuse

10A range unfused

TRANSISTOR hFE TEST (0-1000)

Range

Test Range Test Current

Test Voltage         

NPN & PNP

0-1000

1b=10

µA

Vce=2.8V

4.1.

Accuracy Calculation

Example: Test reading on 200Vdc range is 56.4V. Accuracy is ‘±0.5% of reading ±2 digits’.

Reading ±2 digits = 56.4 ±2 on the last figure i.e. 56.2 to 56.6V.

±0.5% on this range gives 56.2 - 0.5% to 56.6 + 0.5% or 55.9 to 56.9V. Therefore the actual voltage lies between 55.9 and 56.9V.

Note:

Accuracy is specified for a period of one year after calibration and at 18

O

C to 28

O

C with a relative humidity of 80%.

5. MAINTENANCE

p  

WARNING!

Before attempting to open the case, ensure that test leads have been disconnected from measurement circuits to avoid electric shock hazard.

5.1.

For continued protection against fire, replace fuse only with another of the same rating (F200mA/250V Quick Acting). Fuse is located under the circuit board. To 

gain access to the fuse, remove the two screws from the rear of the meter. Lift off the rear cover, replace the fuse and re-assemble in reverse order.

5.2.

If ‘BAT’ appears on the LCD display, it indicates that the battery should be replaced. Repeat the steps detailed in section 5.1 but 

replacing the battery (9V PP3) instead of the fuse. 

5.3.

SPARE PARTS

MM19/F

FUSE, 200mA

MM19/L

LEAD SET FOR MULTI-METER MM19

6. DECLARATION OF CONFORMITY

For Jack Sealey Ltd. Sole UK importer of Sealey Professional Tools.

Declaration of Conformity  

We, the sole UK importer, declare that the product listed below is in conformity with the following standards and directives.

The construction file for this product is held by the Manufacturer and may be inspected,

by a national authority, upon request to Jack Sealey Ltd.

DIGITAL MULTIMETER

Model: MM19.V2

89/336/EEC EMC Directive

73/23/EEC Low Voltage Directive

93/68/EEC CE Marking Directive

Signed by Mark Sweetman

20th December 2004

NOTE:

It is our policy to continually improve products and as such we reserve the right to alter data, specifications and component parts without prior notice.

IMPORTANT: 

No liability is accepted for incorrect use of this equipment. 

WARRANTY: 

Guarantee is 12 months from purchase date, proof of which will be required for any claim.

INFORMATION:

For a copy of our latest catalogue and promotions call us on 01284 757525 and leave your full name and address, including postcode.

Sole UK Distributor,           

Sealey Group,

Bury St. Edmunds, Suffolk.

01284 757500

[email protected]

www.sealey.co.uk

01284 703534

e-mail

Web

MM19.V2 - 1 - 201204