background image

Ac voltAGe
Range

Accuracy

Resolution

200V

±1.2% of rdg ± 10 digits

100mV

600V

±1.2% of rdg ± 10 digits

1V

Overload protection: 600Vdc or 600Vac for all  ranges.

Frequency range: 40Hz - 450Hz

response: Average calibrated in rms sine wave.

teMpeRAtuRe
  Range

                                  Accuracy 

Resolution

0°C to 40°C 

1

-50°C to 400°C 

2

400°C to 1000°C 

2

-50°C

to

1000°C

± 5.0%  of rdg  

± 2°C   

± 0.75% of rdg  

± 3°C

± 1.5% of rdg  

± 15°C

1°C

1. Built in temperature sensor within multimeter.

2. Temperature readings with thermocouple.

3.6  

tRANsIstoR testING

3.6.1 

Set the rotary switch to the hFE position.

3.6.2 

Determine whether the transistor to be tested in nPn or PnP type and locate the Emitter, Base and Collector leads. Insert leads of the transistor into the 

 

correct holes in the external transistor testing socket. Connect the black lead to the COM connector and the red lead to the V

mAºChFE input socket.

3.6.3 

The meter will show the approximate hFE value at test conditions of base current 10µA and Vce 2.8V.

3.7  

coNtINuIty test

3.7.1 

Connect the black lead to the COM input socket and the red lead to the V

mAºChFE input socket (the polarity of the red lead is ‘+’).

3.7.2 

Set the rotary switch to the      position and connect the test leads across the two points of the circuit under test. If continuity exists (i.e. resistance less than  

 

100

) the built-in buzzer will sound.

3.8 

MeAsuRING teMpeRAtuRe

3.8.1 

Set the rotary switch to the 

°

C position and the LCD display will show the ambient temperature.

3.8.2   Insert the red lead of the K-type thermocouple plug into the V

mAºChFE input socket on the front panel and the black lead into the COM. now touch the  

 

object to be measured with the thermocouple probe.

   

WARNING!

 To avoid electric shock, ensure that the thermocouple has been removed before changing to another function measurement.

5.  MAINteNANce

4.  specIfIcAtIoN

   

WARNING!

 Before attempting to open the case, ensure that test leads have  

 

been disconnected from measurement circuits to avoid electric shock hazard.

5.1 

To change a fuse.  The fuses are located on the back of the circuit board. To  

 

gain access to the fuse, remove the protective rubber boot and the two screws  

 

from the rear of the meter. Lift off the rear cover, replace the fuse and 

 

re-assemble in reverse order.

   

WARNING!

 

AlWAys

 replace a fuse with one of the correct rating.

 

fuse 1: 500mA/250v

 

fuse 2: 10A/250v

5.2 

If the battery sign         appears on the LCD display, it indicates that the battery  

 

should be replaced. repeat the steps detailed in section 5.1 to remove the rear  

 

cover, replace the battery (9V PP3) and re-assemble in reverse order. 

NOTE:

 It is our policy to continually improve products and as such we reserve the right to alter data, specifications and component parts without prior notice.

IMpoRtANt: 

no liability is accepted for incorrect use of this equipment. 

WARRANty: 

Guarantee is 12 months from purchase date, proof of which will be required for any claim.

INfoRMAtIoN: 

For a copy of our latest catalogue and promotions call us on 01284 757525 and leave your full name and address, including postcode.

4.1 

Accuracy calculation

 

Example: Test reading on 200Vdc range is 56.4V. Accuracy is ±0.8% of reading ±2 digit.

 

reading ±2 digit = 56.4 ±2 on the last figure i.e. 56.2 to 56.6V.

 

±0.8% on this range gives 56.2 - 0.8% to 56.6 + 0.8% or 55.7 to 57.1V. Therefore the actual voltage lies between 55.7 and 57.1V.

note: 

Accuracy is specified for a period of one year after calibration and at 18

°

C to 28

°

C with a relative humidity of 80%.

MM20.V2      Issue No: 2 - 02/07/10

01284  757500

[email protected]

www.sealey.co.uk

01284  703534

sole u.K. distributor 

sealey Group

Bury St. Edmunds 

Suffolk.

W e b

tRANsIstoR hfe test (external transistor test plug)

Range

test Range

test current

test voltage

nPn & PnP

0-1000

1b=10

µA

Vce=2.8V

dc voltAGe
Range

Accuracy

Resolution

200mV ±0.5% of reading ±2 digits

0.1mV

2V

±0.8% of reading ±2 digits

1mV

20V

±0.8% of reading ±2 digits

10mV

200V

±0.8% of reading ±2 digits

0.1V

600V

±1.0% of reading ±2 digits

1V

Input impedance: 10M

 for all ranges.

dc cuRReNt
Range

Accuracy

Resolution

2mA

±1.0% of reading ±2 digits

1µA

20mA

±1.0% of reading ±2 digits

10µA

200mA ±1.2% of reading ±2 digits

0.1mA

10A

±2.0% of reading ±2 digits

10mA

Overload protection: Fuse 1: 500mA/250V 

 

Overload protection: Fuse 2: 10A/250V   

ResIstANce
Range

Accuracy

Resolution

200

±0.8% of reading ±2 digits

0.1

2k

±0.8% of reading ±2 digits

1

20k

±0.8% of reading ±2 digits

10

200k

±0.8% of reading ±2 digits

100

2M

±1.0% of reading ±2 digits

1k

Maximum open circuit voltage: 2.5V

Overload protection: 220Vdc or rms for a maximum 

of 15 seconds for all ranges.

 

environmental protection.

 

recycle unwanted materials instead of disposing of them as  

 

waste. All tools, accessories and packaging should be sorted, 

 

taken to a recycle centre and disposed of in a manner which is 

 

compatible with the environment.

 

 

When the product is no longer required, it must be disposed of in an 

 

environmentally protective way. 

battery Removal

See Section 5.2.

1. 

remove and dispose of according to local authority guidelines.

2. 

 

Отзывы: