4. Maintenance

Ag/AgCl reference cell (SSC) is maintenance free.

After using the tester:

 1). Rinse the reference cell and test probe in fresh water. 

 2). Dry all parts after use and keep them in the box.

 3). Store in a dry, safe place. 

5. Troubleshooting

Problem

Possible solution

No display

Check that the test probe lead and reference 

cell  are  plugged  into  the  correct  place.  The 

battery is dead. Replace the battery.

Reading is 

inaccurate

Check the wire of the reference cell. Replace 

it if there is any damage. Clean the surface of 

Zinc reference cell using fine sand paper.

Any damage Contact Seaguard International for 

replacement.

6. Specifications

Measurement range

-2.500 to +2.500 volts

Measurement accuracy 0.5%
AD resolution

16 bits

Input impedance

>10

8

 Ω

Display

TFT LCD 1.8”, 320x220

Power 

3 AAA Batteries 

Auto shutdown

No button action for 2.5 minutes

Manual shutdown

Press power button >2 seconds

Data stored

Maximum 50 measurement values 

Working temperature

-10°C to 50°C

Dimensions

142 mm x 63 mm x 32 mm

7. For more information

For more information, please check out our website 

seaguardinternational.com

10

Содержание Digital Corrosion Tester PRO

Страница 1: ...USER MANUAL DIGITAL CORROSION TESTER PROFESSIONAL MODEL ...

Страница 2: ... the furnishing performance or use of this document The information contained in this manual is subject to change without notice This product complies with EU directives Danger Please do not use or store this product near explosive materials strong magnetic fields or a combustible gas environment otherwise it may cause a fire or explosion To avoid being shocked do not use this product when the tes...

Страница 3: ...Using the Selection Window 6 2 3 4 Choosing the boat material 6 2 3 5 Change the potential limit 6 2 3 6 Taking measurements 7 2 3 7 Reviewing saved data 7 2 3 8 Clearing all the saved data 8 2 3 9 Resetting to factory settings 8 3 HowtoConnectandMeasure 8 3 1General 8 3 2 Sterndrive 9 3 3 Inboard 9 3 4Outboard 9 3 5Others 9 4 Maintenance 9 5 Troubleshooting 10 6 Specifications 10 7 For more infor...

Страница 4: ...d it will cause corrosion One of the most effective protection methods is to add sacrificial electrodes called anodes to make the hull become a cathode where the current flows from one to the other To measure the actual corrosion process we measure the electrical potential of the hull If the voltage reading is too high it means the hull is not protected If the voltage is too low we say it is over ...

Страница 5: ...Port for reference cell 6 Port for alligator clip 7 Alligator clip 8 Ag AgCl reference cell 9 Selected material indicator 10 Number of stored reading 11 Type of reference cell 12 Power indicator 13 Upper limit 14 Corrosion protection value 15 Lower limit 16 Right function indicator 17 MENU Save Back Update Upd indicator 18 Left function indicator 11 12 13 14 15 16 17 18 9 10 1 2 3 4 Figure 1 Pro M...

Страница 6: ...perform all functions of the tester Using the selection buttons 1 Power button 2 Move up or increase number 3 Right soft key on the display 4 Go to the Selection Screen or middle soft key on the display 5 Left soft key on the display 6 Move down or decrease number To make selections press the buttons on the tester Figure 3 The words displayed along the bottom of each screen are soft keys that repr...

Страница 7: ... Safe Protected Over protection Steel alloys 0 800V 0 800V to 1 050V 1 050V Aluminum alloys 0 900V 0 900V to 1 100V 1 100V Copper alloys 0 500V 0 500V to 0 700V 0 700V Others 0 500V 0 500V to 0 700V 0 700V The symbol means less than the given limit in the negative direction The symbol means greater than the given limit in the positive direction The default setting is steel 2 3 6 Taking measurement...

Страница 8: ...re 8 Update Saved Data 3 How to Connect and Measure WARNING Do not put the reference cell or test probe into water that is contaminated with oil diesel gas or pollutants This can cause contamination of the reference cell and erroneous readings 3 1 General Lower the reference cell into the water so that it is at least a foot below the water surface near to the boat If there is wave action lower the...

Страница 9: ...E 3 4 Outboard 3 5 Others Any other structure that is steel aluminum or copper which is in water can be measured CLIP TO BOLT ON OUTDRIVE CLIP TO GROUND STRAP OR TO COPPER WIRE CONNECTING ITEMS SHOWN 3 2 Sterndrive 3 3 Inboard 9 ...

Страница 10: ...eference cell Replace it if there is any damage Clean the surface of Zinc reference cell using fine sand paper Any damage Contact Seaguard International for replacement 6 Specifications Measurement range 2 500 to 2 500 volts Measurement accuracy 0 5 AD resolution 16 bits Input impedance 108 Ω Display TFT LCD 1 8 320x220 Power 3 AAA Batteries Auto shutdown No button action for 2 5 minutes Manual sh...

Отзывы: