Section 4: Setting Up MicroTSG
SBE 45
25
Autonomous Sampling
The MicroTSG samples data at pre-programmed intervals, defined by
Interval=
, and sends the data to the computer. The MicroTSG does not enter
quiescent (sleep) state between samples.
Examples:
Autonomous Sampling Mode
(user input in bold)
Example 1
:
J1 jumper in Normal position
(pins 2 and 3)
, AutoRun=Y.
Set up to take a sample every 20 seconds. Send status command to verify setup. Send power-off
command after all parameters are entered. Assuming that power is always applied:
(Click Connect on Toolbar to wake up.)
S>
SINGLESAMPLE=N
S>
INTERVAL=20
S>
AUTORUN=Y
S>
DS
(to verify setup)
S>
QS
When ready to begin sampling:
(Click Capture on Toolbar to capture data to a file – program requests file name for data to be stored.)
(Click Connect on Toolbar to wake up – sampling begins automatically.)
When ready to stop sampling and go to sleep:
(Press Enter key to get
S>
prompt.)
S>
STOP
S>
QS
Example 2
:
J1 jumper in Normal position
(pins 2 and 3)
, AutoRun=N.
Set up to take a sample every 20 seconds. Send status command to verify setup. Send power-off
command after all parameters are entered. Assuming that power is always applied:
(Click Connect on Toolbar to wake up.)
S>
SINGLESAMPLE=N
S>
INTERVAL=20
S>
AUTORUN=N
S>
DS
(to verify setup)
S>
QS
When ready to begin sampling:
(Click Capture on Toolbar to capture data to a file – program requests file name for data to be stored.)
(Click Connect on Toolbar to wake up.)
S>
GO
When ready to stop sampling and go to sleep:
(Press Enter key to get
S>
prompt.)
S>
STOP
S>
QS
Example 3
:
J1 jumper in Autopower position
(pins 1 and 2).
Set up to take a sample every 20 seconds. Send status command to verify setup. Remove power
after all parameters are entered.
(Apply power to wake up.)
S>
SINGLESAMPLE=N
S>
INTERVAL=20
S>
AUTORUN=Y
S>
DS
(to verify setup)
(Remove power.)
When ready to begin sampling:
(Click Capture on Toolbar to capture data to a file – program requests file name for data to be stored)
(Apply power to wake up; sampling begins automatically.)
When ready to stop sampling:
(Remove power.)
To change the setup:
(Apply power to wake up; sampling begins automatically. Press Enter key several times to get
S>
prompt.)
S>
STOP
S>
(send desired commands)
(Remove power.)
Notes:
•
After waking the
MicroTSG, you may
need to press the Enter
key several times and
send
Stop
to interrupt
sampling, depending on
how the instrument was
set up the last time it
was used.
•
If using the MicroTSG
with the Interface Box,
set up the MicroTSG to
operate in autonomous
sampling mode:
¾
J1 jumper in
Autopower position
(pins 1 and 2)
¾
AutoRun=Y
¾
SingleSample=N
30
Содержание SBE 45 MicroTSG
Страница 2: ...This page intentionally left blank ...
Страница 4: ...Manual Generation Date 13 February 2012 1 ...
Страница 6: ...This page intentionally left blank 3 ...
Страница 7: ...4 ...
Страница 98: ...CALIBRATION SHEETS SBE 45 Temperature Calibration S N 0402 1 SBE 45 Conductivity Calibration S N 0402 2 95 ...
Страница 112: ...109 ...
Страница 118: ...6 Appendix I Example Conductivity Calibration Sheet 115 ...
Страница 119: ...7 Appendix II Example Temperature Calibration Sheet 116 ...
Страница 129: ...126 ...
Страница 130: ...127 ...
Страница 131: ...128 ...
Страница 132: ...129 ...
Страница 133: ...130 ...
Страница 134: ...131 ...
Страница 135: ...132 ...
Страница 136: ...133 ...
Страница 137: ...134 ...
Страница 138: ...135 ...