Test Bowl Rinse
1.
Activate bowl rinse.
2.
If bowl does not rinse long enough adjust 3-way bleed valve clockwise
3.
If rinses to short turn 3-way bleed valve counter-clockwise
Test Cup Fill
1.
Activate cup fill.
2.
If bowl does not fill enough adjust 3-way bleed valve clockwise
3.
If cup fills too much turn 3-way bleed valve counter-clockwise