![SDI FS100-RM Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/sdi/fs100-rm/fs100-rm_operating-manual_1220404009.webp)
Copyright © SpectraDynamics, Inc. 2010
7
This unit is designed to operate only with the specified voltage on page 4. For
conversion to a different voltage of operation contact SDI.
To operate the unit, locate the power entry module on the rear of the enclosure and
connect the power cord. Plug the unit into an appropriate power outlet. A LED on the
front panel labeled AC POWER will turn on. The 5 MHz or 10 MHz input signal to be
multiplied should be connected to the SMA connector on the front panel labeled INPUT.
The multiplied signals will be available at the second set of SMA connectors labeled
OUTPUTS. The input signal level must be greater than +12 dBm, otherwise the
synthesizer will not function properly and noise performance will be seriously degraded.
The multiplying synthesizer is not a linear device therefore the output power does
not vary linearly with input power.
The absolute maximum specifications should be
observed to obtain the optimum performance and ensure reliability.
Absolute Maximum Ratings
Input RF Power
+20 dBm Maximum
Reverse RF Power
+20 dBm Maximum
DC Voltage @ Input
20 VDC Maximum
DC Current @ Input
100 mA Maximum
DC Voltage @ Output
50 VDC Maximum
Storage Temperature
-10 to +75 ºC
Operation Environment
0 to +50 ºC
Rack Mount Chassis
1U H, 19“ W, 14” D
Operation