background image

Figure 2.  Using the 963E Benchtop Air Ionizer 

Maintenance 

Occasional cleaning of the case and of the ionizing 

electrodes are the only routine maintenance procedures 

required. 

CLEANING THE CASE 

Wipe the case with a soft cloth moistened with water. 

If a stronger cleaning solution is required, mild soap 

with water may be used. The use of any other cleaning 

solutions is not recommended. 

CLEANING THE EMITTER POINTS 

NOTE: Disconnect the power to the ionizer whenever 

cleaning the emitter points. 

When the emitter pionts become dirty,  the internal 

circuitry of the ionizer will be automatically adjusted to 

emit an equal amount of positive and negative ions. 

Particulates on the emitter point may, however, inhibit 

ionization to a limited degree. The emitter points are 

located between the fan blades and the rear grill. A jet 

of clean, compressed air can be used to remove dirt 

on emitter points. If a more rigorous cleaning method is 

needed to remove particulate,  clean the points with a 

cotton swab dampened with isopropyl alcohol. Access to 

the points is available through the rear grill. Be careful 

not to damage the points during cleaning. 

Calibration 

The SCS Benchtop Air Ionizer is  factory adjusted to 

provide optimum performance.  Further adjustment in the 

field is not possible.  However, the following instructions 

can be followed to determine whether the Ionizer is 

performing to specifications.  The testing follows the 

procedure  outlined in the standard for Ionization, ANSI/ 

ESD STM3.1. Please refer to this standard for more 

complete information. 

Frequency of recalibration should be based on the 

critical nature of those ESD sensitive items handled and 

the risk of failure for the ESD protective equipment and 

materials. In general, SCS recommends that calibration 

be performed annually. 

Periodically measure the balance and neutralization 

times of the Bencthop Air Ionizer to verify that it is 

performing within specifications. These measurements 

should be taken using a charged plate monitor such as 

the SCS  items  711,  770004  and 770005. 

Calibration should be performed in accordance with 

the ESD Association ionization standard ANSI/ESD 

STM3.1. With the fan speed set to high and a distance 

of 12 inches (30 cm), the neutralization (discharge) time 

of ±1 000V to ±1 00V should be less than 2 seconds,  and 

the  balance should be ±1 SV or better. 

TB-9007 

Page 2 of 4 

02020 DESCO  NDUSTRIES,  NC. 

Employee Owned 

Отзывы:

Похожие инструкции для 963E