SCS 770734 Скачать руководство пользователя страница 2

TB-9105

 Page 2 of 4

© 2020 DESCO INDUSTRIES INC

Employee Owned

SCS 

- 926 JR Industrial Drive, Sanford, NC 27332

East: (919) 718-0000 | West: (909) 627-9634 • Website: 

StaticControl.com

D. Rotary Switch:

 Selects the various pass and fail 

load values needed to verify the monitor's operator test 

circuit.

E.  Body Voltage Pushbutton:

 Loads the selected 

body voltage stimulus onto the Operator Test Lead 

when pressed.

F.  Body Voltage LED:

 Illuminates blue when the Body 

Voltage Pushbutton is pressed and voltage stimulus is 

applied onto the Operator Test Lead.

G. Mat Fail Pushbutton:

 Simulates a MAT FAIL 

condition on the 724 Plus Workstation Monitor when 

pressed.

H. Low Battery LED:

 Illuminates yellow when the 

battery power for the body voltage test circuit is low and 

needs to be replaced.

Operation

Verifying the Operator Resistance Circuit

1.  Connect the 724 Plus Monitor Verification Tester's 

green ground lead to equipment ground. This may 

be done using the included Ground Plug Adapter or 

alligator clip. 

2.  Insert the verification tester's black operator test 

lead into the 724 Workstation Monitor's operator 1 

jack. 

3.  Set the rotary switch to 1.5M LOW. The monitor's 

operator 1 LED should blink yellow, and its audible 

alarm should sound. 

4.  Set the rotary switch to 1.5M PASS. The monitor's 

operator 1 LED should illuminate green, and its 

audible alarm should not sound. 

5.  Set the rotary switch to either 10M PASS or 35M 

PASS, whichever one is appropriate. The monitor's 

operator 1 LED should illuminate green, and its 

audible alarm should not sound. 

6.  Set the rotary switch to either 10M HIGH or 35M 

HIGH, whichever one is appropriate. The monitor's 

operator 1 LED should illuminate red, and its 

audible alarm should sound.

Figure 3. Using the Ground Plug Adapter to ground the 

724 Plus Monitor Verification Tester

Figure 4. Connecting the 724 Plus Monitor Verification 

Tester to the 724 Plus Workstation Monitor's operator 1 

jack

Verifying the Operator Body Voltage Circuit

7.  Set the rotary switch to +2.5V FAIL, and press the 

Body Voltage Pushbutton. The monitor's operator 1 

LED should blink red, and its audible alarm should 

sound. 

8.  Set the rotary switch to +2.5V PASS, and press the 

Body Voltage Pushbutton. The monitor's operator 1 

LED should illuminate green, and its audible alarm 

should not sound. 

9.  Set the rotary switch to -2.5V PASS, and press the 

Body Voltage Pushbutton. The monitor's operator 1 

LED should illuminate green, and its audible alarm 

should not sound. 

10. Set the rotary switch to -2.5V FAIL, and press the 

Body Voltage Pushbutton. The monitor's operator 1 

LED should blink red, and its audible alarm should 

sound. 

11. Disconnect the operator test lead from the monitor's 

operator jack. 

12. Connect the operator test lead to the monitor's 

operator 2 jack. Repeat steps 1-11 to verify the 

monitor's operator 2 resistance and body voltage 

circuitry.

Отзывы: