724 Plus Workstation Monitor

Testing the Operator Circuits

The operator high test limit may be configured using 

the Set switch located on the back of the 724 Plus 

Workstation Monitor. The test limit may be set to either 

10 megohms or 35 megohms. Take note of the test limit 

setting before following the instructions listed below.
The Verification Tester’s ground clip toggle switch is 

not used in the verification of the 724 Plus Workstation 

Monitor. It may be set to any position.
1.  Insert the Verification Tester’s stereo plug into the 

monitor’s OPERATOR 1 jack. 

2.  Select “1.5M LOW” with the operator rotary switch.  

The OPERATOR 1 LED should blink yellow, and the 

audible alarm should sound. 

3.  Select “1.5M PASS” with the operator rotary switch. 

The OPERATOR 1 LED should illuminate green, 

and its audible alarm should not sound. 

4.  Select either “10M PASS” or “35M PASS”, 

whichever one is appropriate, with the operator 

rotary switch. The monitor’s OPERATOR 1 LED 

should illuminate green, and its audible alarm 

should not sound. 

5.  Select either “10M HIGH” or “35M HIGH”, whichever 

one is appropriate, with the operator rotary switch. 

The monitor’s OPERATOR 1 LED should illuminate 

red, and its audible alarm should sound. 

6.  Repeat steps 1-5 for the OPERATOR 2 jack on the 

operator remote.

Testing the Mat Circuit

Equipment Needed

•  Resistance Decade Box 

(1 megohm to 5 megohm range, ±1% tolerance)

•  2 Test Leads for the Resistance Decade Box

•  1 Alligator Clip 
1.  Connect the two test leads to the resistance decade 

box. 

2.  Connect one of the test leads to equipment ground. 
3.  Disconnect the 724 Plus Workstation Monitor’s mat 

monitor cord from its worksurface mat. The mat LED 

should illuminate red, and the audible alarm should 

sound. 

4.  Use an alligator clip to connect the second test lead 

from the resistance decade box to the metal snap on 

the mat monitor cord. 

5.  Set the resistance decade box to the values shown 

in the following table. The 724 Plus Workstation 

Monitor’s MAT LED and audible alarm should 

behave as described in the table.  

Load 

Resistance

Mat LED Audible 

Alarm

3.1 megohms

Green

Off

4.3 megohms

Red

On

Specifications

Operating Temperature 50 to 95°F (10 to 35°C)
Environmental 

Requirements

Indoor use only at altitudes 

less than 6500 ft. (2 km)
Maximum relative humidity 

of 80% up to 85°F (30°C) 

decreasing linearly to 50% @ 

85°F (30°C)

Dimensions

3.8" L x 2.4" W x .9" H

(97 mm x 61 mm x 23 mm)

Weight

0.2 lbs. (0.1 kg)

Country of Origin

United States of America

Operator Resistance Values:

Setting

Nominal 

Resistance

% Tolerance of 

Nominal Resistance

1.5M LOW

1.33 Megohms

±2%

1.5M PASS

1.69 Megohms

±2%

10M PASS

8.45 Megohms

±2%

10M HIGH

11.5 Megohms

±2%

35M PASS

29.4 Megohms

±2%

35M HIGH

40.2 Megohms

±2%

Ground Clip Resistance Values: 

Setting

Nominal 

Resistance

% Tolerance of 

Nominal Resistance

SHORT

< 1 Ohm

±2%

10M PASS

4.99 Megohms

±2%

10M HIGH

11.5 Megohms

±2%

These resistance values may be verified using a digital 

ohmmeter. Connect the ohmmeter’s test leads across 

the Verification Tester’s stereo plug. If any value is out 

of specification, the Verification Tester must be returned 

to the manufacturer for repair.