© 2016 DESCO INDUSTRIES, INC. 

Employee Owned 

 

4.    A “PASS” test result is indicated by illumination 

of the green LED. A “FAIL LOW”

 test result is 

indicated by illumination of the red LED. A 

“FAIL

 

HIGH” test result is indicated by illumination of the 

yellow LED. 

If your ESD test fails, check your wrist strap and 

ensure that it is being worn correctly and/or needs 

to be replaced. 

NOTE 

1: The Combo Tester may briefly display “FAIL 

HIGH” once the test button is released at the conclusion 

of a test. Residual voltage in the circuit’s capacitors 

causes the test to continue for a split second after the 

button is released. 

NOTE 2: Failures may be caused by dry skin or minimal 

sweat layer. For wrist straps, try using an approved 

dissipative hand lotion such as 

Menda Reztore™ ESD

 

Hand Lotion 

to your wrist prior to use. 

 

FOOTWEAR TEST 

1.  Toggle the rocker switch to the left toward FOOT 

PLATE. 

2.  While wearing ESD footwear, place one foot on 

the foot plate, and lift the second foot off the floor. 

Make sure there is no wrist cord plugged into the 

WRIST STRAP jack. 

3.  Push down and hold the test switch until the test 

results are displayed.  Ensure that your finger also 

makes contact with the metal ring around the 

pushbutton when performing a test. 

DO NOT touch any other metal while performing 

your test as this will affect your results. 

4. 

A “PASS”

 test result is indicated by illumination 

of the green LED. A “FAIL LOW” test result is 

indicated by illumination of the red LED. A 

“FAIL

 

HIGH” test result is indicated by illumination of the 

yellow LED. 

If your ESD test fails, check your footwear and 

ensure that it is being worn correctly and/or needs 

to be replaced. 

NOTE 1: The Combo Tester may briefly display “FAIL 

HIGH” once the test button is released at the conclusion 

of a test. Residual voltage in the circuit’s capacitors 

causes the test to continue for a split second after the 

button is released. 

NOTE 2: Failures may be caused by dry skin or minimal 

sweat layer. Footwear test results can be improved 

by taking a short walk to build a sweat layer for better 

conductivity. 

5.  Repeat steps 2-4 for the second  foot. 

 

Maintenance 

The Combo Tester requires little maintenance. There 

are no user serviceable parts. If your meter requires 

service beyond cleaning the test switch or foot plate, 

please contact the factory. 

Use a minimum of 80% isopropyl alcohol to clean the 

test switch. Other cleaners are susceptible to leaving 

residue. 

BATTERY REPLACMENT 

The low battery indicator flashes briefly every time 

the unit is used. Replace the battery once the 

indicator becomes constant during testing.  Open 

the compartment located on the back of the tester to 

replace the battery. The tester uses one 9V alkaline 

battery. 

 
 
 
 
 
 
 
 
 
 
 
 
 

 

Figure 12. Replacing the battery 

 
 
 
 

Figure 11. Performing a footwear test with the 

770031 

Combo Tester with Stand and Foot Plate 

 

Texas Technologies, Inc. 

3600 W. Whitestone Blvd. Cedar Park, TX 78613

 

512.267.0100 

 

https://www.texastechnologies.com/static-control-2/testers-meters/ 

E-Mail 

 

[email protected]

Отзывы: