1.Setting of the delayed time of turning on the equipment
This equipment can achieve time delay switch on. When the paramet
er of time delay is not “0”, the
equipment can switch on automatically after time delay.
Hold on the “Change/Confirm” button for 3 seconds, screen displays “CIP”, a second later “0”, press the
“Increase”button to input the password “2”, press the “Change/Confirm” button again and the screen will
display the delay time to switch on the equipment, the symbol “ddtl”and number displays alternately on
the screen(minute,one decimal), press the “Increase”or “Decrease”button to set the parameter, press the
“Change/Confirm” button to confirm the change. Press the “Operating”button at this time, the yellow
indicator HOLD turns on and the preset time begin to count down. When time reduce to “0”, indicator
HOLD turns off, operation begins, the red operating indicator RUN turns on(Under programme control
mode, operation of programme segment begins.).
Simplified instruction on the setting of time delay switch on
2.Setting of the temperature exceeding alarm
Hold on the “Change/Confirm” button for 3 seconds, screen displays “CIP”, a second later “0”, press the
“Increase”button to input the password “3”,press the “Change/Confirm” button again, screen displays
symbol and number alternately:
1)Prnt( the use of interface of printer)
2)ALH(upper limit alarm value,
℃
, one decimal) When the real temperature reaches this value, the
equipment cut off the heating output automatically thus stop the operation. Press the
“
Increase
”
or
“
Decrease
”
button to change the ALH value, then press the
“
Ch
ange/Confirm” button again to confirm
the change.
Attention!
When the parameter of time delay is not “0”, the equipment can
switch on automatically after time delay.
Special Attention:
The upper limit alarm value must be larger than the preset temperature.