![SciQuip Oven-110S Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/sciquip/oven-110s/oven-110s_user-manual_1217847006.webp)
7.Setting for program control mode
6.Setting of parameter under the constant temperature control mode
1.Set temperature&running time
1.Press the SET key and enter the password
3
. Press the SET key again to enter into the temperature
setting function. Press the Increase or Decrease key to set the desired temperature. Press the Set key to
enter into the timer setting function. Press the Increase or Decrease key to set the desired running time.
Press the Set key to confirm the setting.
3
SET
SET
SET
SET
SET
SET
Set temperature
SET
SET
Change/Confirm
button
Hold on for 3 seconds
Increase
button
Change/Confirm
button
Increase
button
Decrease
button
Set running time
Change/Confirm
button
Increase
button
Decrease
button
Change/Confirm
button
Special attention:
The unit has a timer range of 1~999minutes. When the
time parameter is setted as
0
, the unit will keep
running constantly.
Simplified instruction on the setting of constant temperature control mode
The program can be set up to 10 temperature control segments; one ramp step and one hold step for each
segment, and can be run up to 99 times in cycle.
1. Stop the current operation by pressing the "R/H/S" button first; the indicator of "Run" goes off.
2. Press "Set", then press "Increase" and enter code "2", press "Set" again to enter the page of "Segment 1
Ramp Time" (RT). Then press 'Increase/Decrease' button to enter the desired value, and press "Set" again
to confirm. The "Ramp End Temperature" (RET) and "Hold Time" (HT) of each segment can be set in the
same way. (Each segment contains 3 parameter settings, "RT", "RET", "HT")
6.Setting of parameter under the constant temperature control mode
3. For the program with less than 10 segments, user should set a terminating point for the whole
program; simply press "Decrease" to set the "RT" of the followed idle segment as "end", then the
whole program terminates there. Otherwise, set it as "cycle", if repeat running of the whole
program is needed. For example, for a program with only two segments, after finishing setting the
"Segment 2, Hold Time", user should press "Set" to enter the page of "Segment 3, Ramp Time";
then press "Decrease" and set it as "End", if only once operation is needed, alternatively, set as
"Cycle" if multiple repeat operation.
4. Keep pressing "set" to skip the pages of the rest idle segments settings, and enter the page of
"hold segment deviation", enter the deviation temperature limit.
"Hold Segment Deviation (HSD)" is the temperature deviation limit for the condition of entering
the "Hold" step.
4.1. If "HSD" is set as "0", the program enters the "Hold" step on time priority, which means no
matter if the actual temperature reaches at "RET", the "Hold" step will be triggered as soon as the
"RT" dues.
4.2. If "HSD" is set as "N", (N>0), then the program will enters the "Hold" step on temperature
priority, which means the "Hold" step will only be triggered when the actual temperature reaches
at RET
¡À
N
, no matter if the "RT" run out.
7
Start Temperature (
)
Ramp Time (RT) (min)
Ramp End Temperature (RET) ( )
Hold Time (HT) (min)
Segment
30
1
5,(R1)
10,(H1)
-
2
5,(R2)
10,(H2)
Case A: HSD=0
6