SciQuip Oven-110S Скачать руководство пользователя страница 6

7.Setting for program control mode

6.Setting of  parameter under the constant temperature control mode

1.Set  temperature&running time 

1.Press the SET key and enter the password 

3

. Press the SET key again to enter into the temperature 

setting function. Press the Increase or Decrease key to set the desired temperature. Press the Set key to 

enter into the timer setting function. Press the Increase or Decrease key to set the desired running time. 

Press the Set key to confirm the setting.

3

SET

SET

SET

SET

SET

SET

Set   temperature

SET

SET

Change/Confirm

 button

Hold on for 3 seconds

Increase

 

button

Change/Confirm

 button

Increase

 

button

Decrease

 

button

Set   running time

Change/Confirm

 button

Increase

 

button

Decrease

 

button

Change/Confirm

 button

Special attention:

The unit has a timer range of 1~999minutes. When the 
time parameter is setted as 

0

, the unit will keep 

running constantly. 

Simplified  instruction on the setting of constant temperature control mode

The program can be set up to 10 temperature control segments; one ramp step and one hold step for each 

segment, and can be run up to 99 times in cycle.

1. Stop the current operation by pressing the "R/H/S" button first; the indicator of "Run" goes off.

2. Press "Set", then press "Increase" and enter code "2", press "Set" again to enter the page of "Segment 1 

Ramp Time" (RT). Then press 'Increase/Decrease' button to enter the desired value, and press "Set" again 

to confirm. The "Ramp End Temperature" (RET) and "Hold Time" (HT) of each segment can be set in the 

same way. (Each segment contains 3 parameter settings, "RT", "RET", "HT") 

6.Setting of  parameter under the constant temperature control mode

3. For the program with less than 10 segments, user should set a terminating point for the whole 

program; simply press "Decrease" to set the "RT" of the followed idle segment as "end", then the 

whole program terminates there. Otherwise, set it as "cycle", if repeat running of the whole 

program is needed. For example, for a program with only two segments, after finishing setting the 

"Segment 2, Hold Time", user should press "Set" to enter the page of "Segment 3, Ramp Time"; 

then press "Decrease" and set it as "End", if only once operation is needed, alternatively, set as 

"Cycle" if multiple repeat operation.

4. Keep pressing "set" to skip the pages of the rest idle segments settings, and enter the page of 

"hold segment deviation", enter the deviation temperature limit.

"Hold Segment Deviation (HSD)" is the temperature deviation limit for the condition of entering 

the "Hold" step.

4.1. If "HSD" is set as "0", the program enters the "Hold" step on time priority, which means no 

matter if the actual temperature reaches at "RET", the "Hold" step will be triggered as soon as the 

"RT" dues. 

4.2. If "HSD" is set as "N", (N>0), then the program will enters the "Hold" step on temperature 

priority, which means the "Hold" step will only be triggered when the actual temperature reaches 

at RET

¡À

N

, no matter if the "RT" run out.

7

Start Temperature (

)

Ramp Time (RT) (min)

Ramp End Temperature (RET) ( )

Hold Time (HT) (min)

Segment

30

1

5,(R1)

10,(H1)

-

2

5,(R2)

10,(H2)

Case A: HSD=0     

6

Содержание Oven-110S

Страница 1: ...This Manual applies to Please read thisGuide attentively beforeusing and be careful of keepingthe Manual incase of contingentuse User s Manual Oven S Series Blast Air Oven...

Страница 2: ...ire cause by electric leakage 9 5 When moving the equipment Moving the equipment connected with power supply will probably result in electric shock Danger Warnings against likely injury or death Pleas...

Страница 3: ...ration clock indicating poweroff recovery parameter recalling and encryption temperature indicating and correcting RS 232 interface and built in microprinter 4 Monitoring timer independent exceeding g...

Страница 4: ...eating indicator Actual Temperature indicator Time indicator Operating status indicator Setted Temperature Indicator Status keeping indicator Refrigeration indicator Operating button R H S R H S SET S...

Страница 5: ...status indicator When the equipment is well functioned this indicator will be light Status keeping indicator Under program control mode when the equipment is in the state of temperature keeping this...

Страница 6: ...nfirm The Ramp End Temperature RET and Hold Time HT of each segment can be set in the same way Each segment contains 3 parameter settings RT RET HT 6 Setting of parameter under the constant temperatur...

Страница 7: ...ram press the R H S and hold for 3 seconds to quit the existing program first the Run indicators goes off and press R H S again to restart the whole program Case B HSD 0 1 Attention 2 SET SET SET SET...

Страница 8: ...2 Setting of the clock Press the SET key and enter the password 4 The clock time can be changed when deviation between clock time and standard time occurs Press the SET key and enter the password 4 Pr...

Страница 9: ...igned programme the operating indicator RUN turns on 2 When the equipment is in the state of temperature control or programme control the heating indicator lights or twinkles means the heating is conn...

Страница 10: ...arameter is unreasonable Refer to the operating procedure and change or re set parameter 3 Malfunction of heating system occurs Notify the manufacturer for repair service Real temperature is lower tha...

Отзывы: