background image

 

 

Rotary  Evaporator

 

1.

 

Safety Instructions 

 

Read  the  operating  instructions  in  full  before  starting 

up and follow the safety instructions. 

 

Keep  the  operating  instructions  in  a  place  where  they 

can be accessed by everyone. 

 

Ensure that only trained staff work with the appliance. 

 

Follow the safety instructions, guidelines, occupational 

health  and safety  and accident prevention regulations. 

When working under a vacuum in particular! 

 

Wear  your  personal  protective  equipment  in 

accordance with the hazard category of the medium to 

be processed. Otherwise there is a risk of: 

- splashing liquids, 

-  body  parts,  hair,  clothing  and  jewellery  getting 

caught, 

- injury as a result of glass breakage. 

 

CAUTION!  Inhalation  of  or  contact  with  media  such 

as poisonous liquids, gases, spray mist, vapours, dusts 

or  biological  and  microbiological  materials  can  be 

hazardous to user. 

 

Set up the device in a spacious area on an even, stable, 

clean, non-slip, dry and fireproof surface. 

 

Ensure  that  there  is  sufficient  space  above  the  device 

as  the  glass  assembly  may  exceed  the  height  of  the 

device. 

 

Prior to each use, always check the device, accessories 

and  especially  the  glass  parts  for  damage.  Do not  use 

damaged components. 

 

Ensure that the glass assembly is tension-free! Danger 

of cracking as a result of: 

- stress due to incorrect assembly, 

- external mechanical hazards, 

- local temperature peaks. 

 

Ensure  that  the  stand  does  not  start  to  move  due  to 

vibrations respectively unbelance. 

 

Beware of hazards due to: 

- flammable materials, 

- combustible media with a low boiling temperature, 

- glass breakage. 

 

CAUTION! Only process and heat up media that has a 

flash  point  higher  than  the  adjusted  safe  temperature 

limit of the heating bath that has been set. 

 

The  safe  temperature  limit  of  the  heating  bath  must 

always be set to at least 25 °C lower than the fire point 

of the media used. 

Содержание SCI100-Pro

Страница 1: ...rator User Manual LCD Digital Rotary Evaporator Please read the User Manual carefully before use and follow all operating and safety instructions Technical specifications and outline are subject to ch...

Страница 2: ...ion 6 3 1 Receiving Inspection 6 3 2 Listing of Items 6 4 Control and Display 6 4 1 Control elements 6 4 2 Display 8 5 Installation and Operation 9 5 1 Installation 9 5 2 Operation 13 6 Faults 14 7 Ma...

Страница 3: ...or software Methods and procedures adopted to resolve the problems Your contact information Warranty This instrument is warranted to be free from defects in materials and workmanship under normal use...

Страница 4: ...materials can be hazardous to user Set up the device in a spacious area on an even stable clean non slip dry and fireproof surface Ensure that there is sufficient space above the device as the glass a...

Страница 5: ...r a compa rable protective device Adapt the quantity and the type of distilland to the size of the distillation equipment The cooler must work properly Monitor the coolant flow rate at the cooler outl...

Страница 6: ...ust gases indicate that flask content is beginning to decompose Switch off heating immediately Use the lifting mechanism to lift the evaporator piston out of the heating bath Evacuate the danger zone...

Страница 7: ...rotect the appliance and accessories from bumps and impacts The appliance may only be opened by experts 2 Proper Use The instrument is designed for distilling in schools laboratories or factories Toge...

Страница 8: ...damage to the system please do not plug it into the power line 2 Loosen transportation lock before Power ON 3 2 Listing of Items The package includes the following items Items Qty Main unit 1 Heating...

Страница 9: ...ater RT 90 press the key for 3s to switch the heating mode Timer key Timer Push the key and then rotate control knob to set the rated time in the range from 1 to 999min Int key Int Push the key and th...

Страница 10: ...rotating function is switched ON Temp Display when the heating function is switched ON Timer Display when timer function is switched ON Int Display when Int function is switched ON Speed display area...

Страница 11: ...heating mode to oil press the temperature key for 3s the temperature display OIL Figure 6 Figure 6 When select the oil mode the temperature display as Figure 7 Figure 7 The same way press the temperat...

Страница 12: ...nd position The distance is 150mm Connect the device to the power supply using power cable b Connect the heating bath to the device c Figure 7 Remove the clamping device for the angle setting of the r...

Страница 13: ...it clockwise by 60 The plastic screw nut f helps loosen tight fitting piston ground in connections Figure 9 Hold the tight fitting evaporator piston and counter clockwise e and then turn the plastic s...

Страница 14: ...locking spring i to condenser in turn Tighten the cap nut hand screwed h Start the device at 200 rpm for 2 min Then tighten the cap nut firmly Figure 12 Figure 13 Installing the vertical glassware co...

Страница 15: ...15 Figure 15 5 2 Operation Make sure the required operating voltage and power supply voltage match Ensure the socket must be properly grounded Plug in the power cable ensure the power is on and begin...

Страница 16: ...0 Press the control knob switch ON If these operations above are normal the device is ready to operate If not the device may be damaged during transportation please contact manufacturer supplier for...

Страница 17: ...de Before using other method for cleaning or decontamination the user must ascertain with the manufacturer that this method will not harm the instrument Wear the proper protective gloves during cleani...

Страница 18: ...igital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment Th...

Страница 19: ...tion Yes Temperature Range C RT 180 Heat Control Accuracy C 1 Temperature Display LCD Heat Output W 1300 Lift Motor Stroke mm 150 Timer min 1 999 Overall Dimensions D W H mm 465 457 583 Weight kg 15 P...

Страница 20: ...SCILOGEX LLC 1275 Cromwell Ave Suite C6 Rocky Hill CT 06067 USA Tel 1 860 436 9221 Fax 1 860 436 9745 info scilogex com www scilogex com...

Отзывы: