Scilogex MS-H-Pro Скачать руководство пользователя страница 11

9

MS-(H)-Pro Digital(Hotplate)Magnetic Stirrer

Operation mode B 

    MS-Pro maintains the state of the heating and stirring 

functions before the last switching off (mains OFF), as well 

as the data set last can be changed.

    Remote control can be used. 

    Stirring bar breakaway monitoring can

t be set in this 

mode.

Note:

Mode changing and parameter setting is forbidden 

when the instrument is running.

 Switch OFF 

Stir

and 

Heat

control knob 

in advance before change mode.

8  Function: Heating(MS-H-Pro)

The device is controlled by Digital Temperature Control 

Technology ,which has two separate safe circuit.  The 

hotplate is kept at a constant temperature by a digital control 

circuit. The hotplate temperature can also be monitored from 

a separate, adjustable safety circuit. The two temperature 

sensors (Pt1000) internal for temperature control are built 

into the hotplate.The single external Pt1000 can monitor the 

temperature of sample.

If external temperature sensors needed, it must be plugged in 

before the device is switched ON.

•    Ensure the external temperature sensors connecting 

firmly before heating.

•    Set the temperature via turnning the temperature control 

knob slowly to the ideal scale .

•    When the heating function is switched on, the LED 

Heat

 will light up and the LCD will display the real 

temperature.

•    The rated temperature will be displayed on the right-hand 

side of the LCD as well as Temp and 

°

C characters.

•    The heating function is switched on or off via pushing the 
heat knob.

8.1 Working with external temperature sensor

Figure 5

Содержание MS-H-Pro

Страница 1: ...Digital Hotplate Magnetic Stirrer MS Pro Digital Magnetic Stirrer Please read the User Manual and the related Video of this instrument on our website carefully before use and follow all operating and...

Страница 2: ......

Страница 3: ...tion Mode of MS H Pro 8 7 2 Operation Mode of MS Pro 8 8 Function Heating MS H Pro 9 8 1 Working with External Temperature Sensor 9 8 2 Residual Heat Warning HOT 10 8 3 Setting the Safety Temperature...

Страница 4: ...rial Number on the rear panel Description of problem i e hardware or software Methods and procedures adopted to resolve the problems Your contact information Warranty You have purchased a Scilogex ins...

Страница 5: ...late Temperature must always be set to at least 25 lower than the fire point of the media used Beware of hazards due to Flammable materials or media with a low boiling temperature Overfilling of media...

Страница 6: ...tip of the measuring sensor must be at least 5 mm from the vessel bottom a distance of 10 mm is ideal The instrument can only be disconnected from the main power supply by pulling out the mains plug o...

Страница 7: ...required operating voltage and power supply voltage match Ensure the socket must be earthed reliably Ensure the power be off Plug in the power cable ensure the power be on and begin initializing Add t...

Страница 8: ...FF via push the knob Temperature control knob Heat MS H Pro Set the rated temperature The function heating is switched ON or OFF via push the knob Mode Key Mode Shift among the three operating modes A...

Страница 9: ...se of error happening Rated value Real value Display value in case of heating or stirring function switching ON Table 3 Note If both heating and stirring functions have been started at the same time t...

Страница 10: ...sidual temperature warning safe temperature limit and stirring bar breakaway monitoring can be set in this mode Operation mode B MS H Pro has heating and stirring functions Users can control on the fr...

Страница 11: ...cuit The hotplate temperature can also be monitored from a separate adjustable safety circuit The two temperature sensors Pt1000 internal for temperature control are built into the hotplate The single...

Страница 12: ...s continuously the function setting key till Set disappear The residual heat warning function has been switched on or off In order to prevent the risk of burns from a hotplate MS H Pro has a residual...

Страница 13: ...n set Breakaway can occur when stirring at high speeds when large stirring bars are being driven at speeds close to the maximum magnetically torque or sample viscosity changes Using MS H Pro stirring...

Страница 14: ...l status standard RS232 C corresponding with DIN66259 Part 1 applies Transmission method Asynchronous signal transmission in start stop operation Mode of transmission Fully Duplex 1 start bit 7 charac...

Страница 15: ...ment when cleaning Unplug the power line when cleaning Only use cleanser that we advised as below Dyes Isopropyl alcohol Construction materials Water containing tenside isopropyl alcohol Cosmetics Wat...

Страница 16: ...cy Hz 50 60 Power W 550 MS H Pro 50 MS Pro Stirring point position quantity 1 Max stirring quantity H2O L 20 Max magnetic bar mm 80 Motor type External rotor brushless motor Max power input of motor W...

Страница 17: ...S H Pro Digital Hotplate Magnetic Stirrer Porcelain enamel USA plug 110V 50Hz 60Hz 811123039999 MS H Pro Digital Hotplate Magnetic Stirrer Stainless Steel Cn plug 220 50Hz 60Hz 811123049999 MS H Pro D...

Страница 18: ...135 5 Quarter 40 ml reaction vessel 18900048 MS 135 6 Quarter 8ml reaction vessel 18900049 MS 135 7 Quarter 16 ml reaction vessel 18900006 Stirring bars 10mm x 6mm 1pcs pk 18900007 Stirring bars 15mm...

Страница 19: ...circuit board 2 Hole for external temperature sensor holder 12 Direct current Motor without brush 3 Mains switch 13 Turntable 4 Housing 14 Internal temperature sensors 5 Ventilation 15 Permanent magn...

Страница 20: ...SCILOGEX LLC 500 Four Rod Road Suite 122 Berlin CT 06037 USA Tel 1 860 828 5614 Fax 1 860 828 5389 E mail info scilogex com Website Http www scilogex com...

Отзывы: