USER MANUAL
2
Make sure to prepare necessary safety measures before using samples that are toxic, radioactive or
contaminated with pathogenic micro-organisms at your own responsibility.
If the instrument, rotor and/or accessories that has been contaminated by solutions with toxic, radioactive
or pathogenic materials, clean it according to the decontamination procedure that you are specified.
If you require services at site, please sterilize and decontaminate it in advance, and then notice the service
center involved in the details of the particular materials.
Do not handle the power cord or turn on or off the POWER switch with wet hands to void electrical
shocks.
For safety purposes, do not enter within 30cm around this centrifuge while it is in operation.
While the rotor is rotating, never forcedly release the door lock.
Unauthorized repairs, disassembly, and other services to the centrifuge except by our service center are
strictly prohibited.
CAUTION
This centrifuge must be located on one firm and level table.
Make sure the centrifuge is horizontal before running.
Make sure the angle between the door and cover is greater than 70 degrees when open the door.
Be careful not put your fingers or hands between the door and cover when the door off.
Do not move or relocate this centrifuge while it is running.
If fluid spills in the rotor chamber, please promptly clean and dry with a dry cloth to avoid sample
contamination.
Ensure to remove any objects and fragments of the tubes dropped inside the rotor chamber before running
this centrifuge.
Cautions on rotors
(1) Always check for corrosion and damages on the rotor surface before using it. Do not use the rotor if an
abnormality is found.
(2) Do not set the centrifuge speed beyond the allowable minimum speed of the rotor kits (rotor or adapters).
Make sure to run it below the allowable minimum speed.
(3) Do not exceed the allowable imbalance.
(4) Use the rotor and tubes within their actual capacities.
(5) If the rotor is attached with a lid, ensure it is tightened before operation.
If any abnormal condition occurs during operation, please stop it immediately and contact our service
center. Notify the service center is a warning code if displayed.
Vibrations are likely to damage the centrifuge, contact our service center if abnormality observed.