4. The meter will automatically recognize the
buffer value, which will show on the screen.
5. Wait up to 10 seconds and "SA" will show
on the screen indicating the calibration at
that point is complete.
6. The meter will briefly display End and return
to normal mode.
7. Repeat steps 2-5 for the pH 4 and 10
buffers.
CUSTOM CALIBRATION
During calibration mode, you can adjust the
buffer values to non-standard values. For
example,
4.00 pH may be adjusted from
3.50 to 4.50.
7.00 pH may be adjusted from
6.50 to 7.50.
10.00 pH may be adjusted from
9.50 to 10.50.
1. Place the electrode in the non-standard
buffer and press the CAL button. The meter
will display 4, 7, or 10 depending on which
is closest.
2. Press and hold the Hold button. The
number on the screen will begin to change in a
positive direction.