scigiene SMARTMETER Скачать руководство пользователя страница 4

SmartMeter 

 

4/4 

L(.(2 

(

L

ow 

C

alibration Channel 2)

 

INPUT OFFSET CALIBRATION

: Adjusts the lowest limit of the input 

signal  indication  range  (offset).  In  order  to  change  the  indicated 
value, it may be required to press 

 

 or 

 

 several times. 

k(.(2 

(

High

 

C

alibration Channel 2)

 

INPUT  SPAN  CALIBRATION

:  Adjusts  the  highest  limit  of  the input 

signal indication range (span). In order to change the indicated value, 
it may be required to press 

 

 or 

 

 several times. 

5

 

CALIBRATION 

The meter is factory calibrated, and recalibration by non-experienced 
professionals  is  not  recommended.  If  required,  calibration  must  be 
performed by expert professionals, who must proceed as it follows: 

a)

  Configure  the  type  of  input  to  be  calibrated:  K  thermocouple  or 

Pt100 or 0-50 mV. 

 

Note

:  By  calibrating  K  thermocouple,  all  the  remaining 

thermocouples  will  be  automatically  calibrated.  Pt100  and  0-50 
mV required their own calibration processes. 

b)

  Access the Calibration cycle. 

c)

  Apply  to  the  input  channel  a  signal  corresponding  to  a  known 

indication,  and  slightly  over  the  lowest  limit  of  input  type 
indication. 

d)

  To  access  parameter 

L(.(1

 (or 

L(.(2

). Use keys 

 

 and 

 

 

until  the  meter  indicates  the  value  corresponding  to  the  applied 
signal. 

e)

  Apply  to  the  input  channel  a  signal  corresponding  to  a  known 

indication,  and  slightly  below  the  highest  limit  of  input  type 
indication. 

f)

  Access parameter 

k(.(1 

(or 

k(.(2.

 Use keys 

 

 and 

 

 until 

the meter indicates the value corresponding to the applied signal. 

g)

  Repeat 

c

 to 

f

 until no adjustment is further required. 

Note

:  While  calibrating  the  meter,  observe  if  the  Pt100  excitation 

current  required  by  the  calibrator  used  is  compliant  to  the  Pt100 
excitation current used in this instrument. 0.165 mA. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6

 

WARRANTY 

The  manufacturer  products  are  covered  by  a  12-month  warranty 
provided the purchaser presents the sales receipt and the following 
conditions are met: 

1.

 

Products are covered for one year from the original date of 
purchase. Please retain the dated sales receipt as evidence 
of  the  date  of  purchase.  You  will  need  it  for  any  warranty 
service.

 

2.

 

Within  this  period, warranty against defects in material and 
workmanship under normal use is free of charge.

 

3.

 

For  repair,  send  the  product  and  the  sales  receipt  to  our 
address.  Expenses  and  transportation  risks  are  under  the 
purchaser’s responsibility

 

4.

 

This warranty does not cover any damage due to accident, 
misuse, abuse, or negligence. 

 

 