scigiene SMARTMETER Скачать руководство пользователя страница 1

 

 

 

SMARTMETER 

PORTABLE TEMPERATURE METER - INSTRUCTIONS MANUAL – V1.0x

 

1

 

INTRODUCTION 

SmartMeter is a portable temperature meter with high-contrast LCD.  
The  Smart  Meter  housing  enables  industrial  environment  or  bench-
top  operation  and  provides  a  belt  carrying  fixture.  It  supports  2 
temperature  sensors  with  independent  and  concurrent  display  of 
measured  values.  It  provides  the  capability  of  reading  0-50  mV 
electric signals with configurable indication scale. 
Difference between channels, relative measure, minimum, maximum 
and  average  values,  hold  and  alarm  functions  are  also  available.  It 
also features an auto power off resource in case the device is idle, in 
order to extend battery life. 
The device is supplied with a K-type thermocouple with 1.5 m cable. 

1.1

 

UNPACKING 

The  meter  is  supplied  in  proper  packaging  that  offers  suitable 
protection  during  transit  and  storage.  Unpack  with  care.  Upon 
delivery,  make sure that there is no damage to the instrument, and 
that there are no missing items. 
Within  packaging,  you  should  find,  besides  the  manual,  the  meter 
and a K-type thermocouple sensor. 

2

 

SPECIFICATIONS 

Input types and max. indication range 

J thermocouple .......................................................... 

100 to 760 ºC 

K thermocouple....................................................... 

-

 150 to 1.370 ºC 

T thermocouple .......................................................... 

-

 160 to 400 ºC 

E thermocouple............................................................. -90 to 720 ºC 
N thermocouple ...................................................... 

-

 270 to 1,300 ºC 

R thermocouple ........................................................ 

-

 50 to 1,760 ºC 

S thermocouple......................................................... 

-

 50 to 1,760 ºC 

B thermocouple......................................................... 500 to 1,820 ºC 
Pt100 ......................................................................... 

-

 200 to 600 ºC 

Voltage 0-50 mV ................................... 

-

 1,999 to 9,999 (Adjustable) 

All  input  types  are  factory  calibrated.  NBR  12771,  Pt100  NBR 
13773/97 (IEC-751) (385) calibrated thermocouples. 
Accuracy: 

Thermocouples: .............................. 0.25 % F.S. 

 1 ºC 

 

Pt100: ............................................... 0.2 % F.S. 

 1 ºC 

 

0-50 mV: ........................................ 0.2 % F.S. 

 1

 

digit 

Resolution: ...................................................................................... 0.1 ºC 
Sampling rate: ............................................................ 2 timer per second 
Input impedance: ............................... > 10 MΩ (Thermocouples, 50 mV) 
Pt100 excitation current: ............................................................ 0.165 mA 
Power supply: ...................................................................... 9 Vdc battery 
AC / DC adapter  ............................................................................. 6 Vdc 
Estimated autonomy: .................................... 400 hours (alkaline battery) 
Dimensions: ................................................................. 165 x 73 x 36 mm 
Weight: ............................................................................................ 205 g 

Protection

:  .......................................................................................... IP40 

3

 

CONNECTIONS 

Female  input  connectors  suitable  for  RTD  /  TC  compensated 
connectors, as shown below. 

 

Figure 01

 — Meter supported connectors 

Only  two  input  terminals  are  used  to  connect  thermocouples.  Input 
terminals  have  different  size  holes,  which  prevent  inappropriate 
installations. 

 

Figure 02

 — Thermocouple connection 

All three terminals are used for Pt100 (RTD) 

 

Figure 03

 – Pt100 connectors sensors 

The Pt100 internal connections should be made as shown in figure 
below. 

 

Figure 04

 — Internal connections of Pt100 sensors 

For the 0-50 mV signal, installation should comply with the following 
connections: 

 

Figure 05

 — 0-50 signal connection 