9. On the MS tab, click
Edit Parameters
.
10. On the Source/Gas tab, specify values in the fields as required.
11. On the Compound tab, specify values in the fields as required and then click
OK
.
12. Click a device icon and then select the parameters for the device.
13. Add any additional periods and experiments. Refer to
14. Click
File > Save.
Add an Experiment
1. In the Acquisition method pane, on the period where the experiment is to be added, right-click and then click
Add experiment
.
Figure 7-1 Add Experiment
An experiment is added below the last experiment in the period.
Note:
An experiment cannot be inserted between experiments, IDA criteria, or periods. Users can only add
an experiment at the end of the period.
2. In the MS tab, select the appropriate parameters.
Add a Period
• In the Acquisition method pane, right-click the
Mass Spec
icon, and then click
Add period
.
A period is added below the last period created.