Figure 6-4 Interface Latch Release Buttons
7. Remove the interface components and then put them on a clean, stable surface.
Tip!
Use the shaped foam insert from the package to store the standard interface components.
Install the Interface Components
Use this procedure to install the standard or NanoSpray
®
interface components on the mass spectrometer.
1. Find the six contact pins on the interface components and the corresponding sockets on the mass spectrometer.
These pins and sockets act as keys to prevent the interface components from being installed in the wrong
orientation.
2. Orient the interface components so that the six contact pins align with the sockets when the two retaining
pins are inserted in their clamps. Refer to
.
Operator Guide
NanoSpray
®
Ion Source
RUO-IDV-05-0811-F
54 / 92
Ion Source Maintenance