Figure 5-2 Stop Pumping
Replace the Mobile Phase with a Miscible Solvent
1. Pour 100 mL of the new mobile phase into a 200 mL beaker.
2. Remove the suction filter from the reservoir and then put the filter into the beaker filled with the new mobile
phase in the preceding step.
Figure 5-3 Replace the Mobile Phase
ExionLC
™
100 System
Hardware User Guide
107 / 219
RUO-IDV-05-1839-C
Service and Maintenance