Table D-1 Parameter Optimization for the TurboIonSpray Probe (continued)
Parameters
Typical Values
Operational
Range
Low
Medium
High
Probe vertical
micrometer setting
7 to 10
2 to 5
0 to 2
0 to 13
Probe horizontal
micrometer setting
4 to 6
4 to 6
4 to 6
0 to 10
APCI Probe Parameters
Table D-2 Parameter Optimization for the APCI Probe
Parameter
Typical Value
Operational Range
LC flow rate
1,000 µL/min
200 µL/min to 3,000 µL/min
Ion source gas 1 (nebulizer
gas)
30 psi
0 psi to 90 psi
Gas for the Curtain Gas
interface
20 psi
20 psi to 50 psi
400 ºC
100 ºC to 750 ºC
Nebulizer current
Positive: 3 µA
Negative: –3 µA
Positive: 0 mA to 5 µA
Negative: –5 mA to 0 µA
Nebulizer Current (NC)
Positive: 2 µA
Negative: –2 µA
Positive: 1 mA to 5 µA
Negative: –1 mA to 0 µA
Declustering potential (DP)
Positive: 60 V
Negative: –60 V
Positive: 0 V to 300 V
Negative: –300 V to 0 V
Probe vertical micrometer
setting
4
Scale 0 to 13
3
Temperature values depend on the compound.
Source Parameters and Voltages
System User Guide
RUO-IDV-05-6475-C
SCIEX 3200 Systems
99/114