ENVIRONMENTAL CHAMBER ONLY
To use the Environmental Chamber only, within the
MIXER MODE
, choose the setting
OFF
. All mixing
will become disabled and the
Enviro-Genie
becomes a temperature chamber only. A wire rack
is included to increase the capacity of the chamber.
An additional rack (sold separately) may be stacked
on the first rack to further increase capacity (see
Accessories).
RS-232 INTERFACE
Enviro-Genie
comes equipped with a serial
interface for connection to a PC or other serial
device. This interface can be used to record
machine information (including temperature) for
situations that require verification. It can also be
used to send commands remotely to the
Enviro-
Genie
, either manually or automatically from a
program in order to create profiles. Visit our
website (www.scientificindustries.com) for a sample
visual basic control program for the
Enviro-Genie
.
The serial port is on the side of the unit and is a
standard subminiature 9-pin male connector.
Connection to a PC is via a straight through 9-pin
female-female serial cable (P/N SI-1132).
Baud Rate: 19,200
Data Bits: 8
Stop Bits: 1
Parity: None
All communication is accomplished with ASCII
Uppercase characters with a carriage return as the
terminating character.
CONTROL COMMANDS
TIME:
Mixing and temperature control time.
Enter commanded time as follows: To
enter 7 Hrs. 23 Min. type “Time 723”
followed by the enter key.
TON:
Turn on data output. Data is output
once per second. To turn on data
output, type the following: “TON”
followed by the enter key. The format of
the data out is as follows: Commanded
Temperature,
Time
and
Actual
Temperature.
TOFF:
Turn off data output. To turn off data
output, type the following: “TOFF”
followed by the enter key.
TEMP:
Temperature command in degrees
Celsius. Will accommodate tenths of
degrees.
To
enter
commanded
temperature, type the following: “TEMP
250” followed by the enter key. This will
return 25.0 degrees Celsius on the front
panel. “TEMP 251” followed by the
enter key will return 25.1 degrees
Celsius. The numerical value is the
temperature value in degrees Celsius in
the range from 4
- 75
.
SPEED:
Speed command in RPM. To enter the
commanded speed, type the following:
“SPEED 500” followed by the enter key.
The same ranges for Rock/Rotate (0-35
RPM) and Stirring (0-1000 RPM)
entered on the keypad hold for RS-232
speed command.
MMODE: Mixing mode. To enter mixing mode
type the following: “MMODE 1-3”
followed by the enter key.
1. No Mix mode.
2. Rock/Rotate mode.
3. Stir mode.
TMODE: Timer mode. To enter timer mode type
the following: “TMODE 1-4” followed by
the enter key.
1. Nothing off.
2. Mixing off.
3. Temperature off.
4. Mixing and temperature off.