Functional Checks 

The 

30V- 2A Power Supply

 should regularly be tested to assure proper functioning. 

The  following  test  checks  out  the  power  supplies  performance  and  suggestions  for 
adjusting  specific  values.  The  adjustment  will  only  be  meaningful  if  the  below 
indicated or equivalent instruments are used. Prior to the functional test or adjustment, 
the instrument should be on for at least 30 minutes. 

Measuring Instruments required : 

1.

 

4 ½ Digit DMM; 4011 Handheld or equivalent 

2.

 

20 MHz Oscilloscope; Caddo 802 or equivalent  

3.

 

Rheostat: 100 

Ω

- 2A; 17 

Ω

- 5.5A 

 

Test Procedure : 

0-30V DC output 

1.

 

Check  of  maximum  DC  output  voltage  :

  Set  the  Coarse  and  Fine  knobs  to 

maximum,  the  maximum  DC  output  reading  should  be  between  31V  and  33  V. 
This also can be verified on DMM. 

2.

 

Check  of  minimum  DC  output  voltage  :

  Set  the  Coarse  and  Fine  knobs  to 

minimum,  the  minimum  DC  output  reading  should  be  00.0  Volts,  which  when 
measured on DMM will be approximately 25mV for load resistor of < 5K

Ω

.  

3.

 

Check of minimum current limit setting :

 Set DC output to 10V & short circuit 

the output terminals. Adjust current limit  knob  I max to minimum the reading on 
the display should be < 100mA. 

4.

 

Check of maximum. Output current :

 Set the DC output voltage to 10V. Short-

circuit the output terminals. Adjust the I max current limit knob to maximum, the 
reading on the display, should be between 2.02A and 2.2A. 

5.

 

Check  of  over  load  Indicators  :

  Set  the  instrument  as  in  3.  When  the  output 

terminals are short circuited, CC LED should light up. 

6.

 

Check of residual ripple and noise :

 Connect any load on DC output and adjust 

the I max, to maximum (maximum to 2000mA), and check the ripple and noise on 
oscilloscope, the ripple and noise should not be more than 1mVrms. 

 

Содержание ST4073

Страница 1: ...erating Manual Ver 1 1 An ISO 9001 2000 company 94 101 Electronic Complex Pardesipura Indore 452010 India Tel 91 731 2570301 02 4211100 Fax 91 731 2555643 e mail info scientech bz Website www scientec...

Страница 2: ...ST4073 Scientech Technologies Pvt Ltd 2...

Страница 3: ...is mandatory that service engineers use lead free solder wire and use the soldering irons upto 25 W that reach a temperature of 450 C at the tip as the melting temperature of the unleaded solder is hi...

Страница 4: ...ntry Ground the Instrument This instrument is grounded through the protective earth conductor of the mains cord To avoid electric shock the grounding conductor must be connected to the earth ground Be...

Страница 5: ...for complex analog and digital applications The DC output can be continuously adjusted from 0 30V with coarse and fine controls Current limit is also adjustable from 0 2A Any over loading for adjuste...

Страница 6: ...cient 0 05 5 mV C Ripple Noise 1mVrms Current Limit Adjustable between 100mA to 2A Display 3 digits for voltage and current Accuracy 1 of reading 1 digit Over Range Indication By flashing CC LED Gener...

Страница 7: ...ing knob For the coarse setting of the output voltages 4 I Max adjusting knob For current limit setting adjustment range 2mA 2A 5 Output 0 30 V Output terminals for 4mm banana plugs or cable connectio...

Страница 8: ...n cord without a protective Conductor Warning Any interruption of the protective conductor inside or outside the instrument or disconnection of the protective earth terminal is likely to make the inst...

Страница 9: ...other Power Supply units to increase supply voltage or current respectively The output power of the 30V 2A Power Supply is not activated by switching on the instrument This is for protection of the lo...

Страница 10: ...an be verified on DMM 2 Check of minimum DC output voltage Set the Coarse and Fine knobs to minimum the minimum DC output reading should be 00 0 Volts which when measured on DMM will be approximately...

Страница 11: ...The customer resells the instrument to another party d Any attempt is made to service and modify the instrument 3 The non working of the product is to be communicated to us immediately giving full de...

Отзывы: