SANAO

 

Handpieces

 

- Installation and Commissioning

SD-468 US EN R1. 

SANAO, E-STAT

IS

 and 

Your Infection Control Specialist

 are trademarks of SciCan Ltd.

Calibrate Foot Control

Press “Set up” key.
The display then goes in to set up mode.

Pressing the button starts the “psi/bar” calibration process.

Proceed as follows:

1.  If “MAX” is not already displayed on the “PSI/BAR” button, press “PSI/BAR” once or twice until 

“MAX” appears displayed on the “PSI/BAR” button.

2.  When “MAX” is displayed the unit is ready to activate calibrating the maximum pressure.

3.  Press the foot control ALL THE WAY DOWN and keep holding it down. Press and HOLD “PSI/

BAR” button, (while “MAX” is displayed and foot control is being held all the way down), until signal 
sounds. Maximum pressure has now been calibrated (to equal 100%).

4.  The “PSI/BAR” button should now have “MIN” displayed on it. If it does not, then press “PSI/BAR” 

once or twice until “MIN” appears displayed on the “PSI/BAR” button

5.  When “MIN” is displayed the unit is ready to activate calibrating the minimal pressure.

6.  DO NOT PRESS THE FOOT CONTROL AT ALL, and Press and HOLD “PSI/BAR” button, (while 

“MIN” is displayed), until signal sounds. Minimum pressure has now been calibrated (to equal 0%).

Calibration is finished.

100%

0%

SD-468 US EN R2_QRG TestDrive_Sanao.indd   2

16-08-19   1:56 PM