Operating Instructions EMP 4 Expert
4 . Functional safety
- Checking the output signals with a load resistance of 1 kΩ real
current and voltage
- Checking the frequency
- Checking the pulse width
These technical checks may only be performed by individuals with
appropriate training .
The results must be noted in the medical device book along with the
date and name of the person carrying out the check .
14. Combination
The
EMP 4 Expert
may be used together with all accessories men-
tioned in the chapter „Accessories“
1. Scope of supply
1
EMP 4 Expert
REF 104004
2 Rechargeable battery WA 60-8
REF 104706
1 Charging station AL 1 A
REF 104709
1 Charger P 60-8
REF 104704
1 Electrode cable type 5 .15 (black)
REF 106351
1 Electrode cable type 5 .16 (grey)
REF 106352
1 Electrode cable type 5 .17 (red)
REF 106353
1 Electrode cable type 5 .18 (yellow)
REF 106354
1 Pack self adhering electrodes
REF 283400
50 mm x 50 mm (Packing Unit = 4 electrodes)
2 Packs self adhering electrodes
REF 283600
50 mm x 90 mm (Packing Unit = 2 electrodes)
1 Transport box
REF 450980-0039