schwa-medico Operating Instruction - EMP 2 PRO
Programme parameters
Programme 1, 3, 4, 8, 9, 10, 11:
1. frequency: 20 Hz up to 120 Hz (Fig. 7)
2. pulse width: 70 µs up to 500 µs (Fig. 8)
3. total run time: 1 up to 99 min. (Fig. 12)
Programme 2, 5 (TENS):
channel 1 gate c channel 2 endorphine release
1. frequency channel 1: 80 Hz, 90 Hz, 100 Hz, 110 Hz, 120 Hz (Fig. 9)
2. frequency channel 2: 0,5 Hz, 1 Hz, 2 Hz, 5 Hz, 10 Hz
3. pulse width: 60 µs up to 500 µs (Fig. 8)
4. total run time: 2 up to 99 min. (Fig. 12)
Programme 6:
1. total run time: 1 up to 99 min. (Fig. 12)
Programme 7 :
Epicondylites modulation
1. max. frequency: 1 Hz up to 120 Hz (Fig. 10)
2. min. frequency: 0,5 Hz up to max. frequency –8 (Fig. 11)
3. pulse width: calculated according to the frequency
4. total run time: 1 up to 99 min. (Fig. 12)
Programme 12, 13, 14, 15, 16, 18, 19:
muscle stimulation
1. frequency: 20 Hz up to 120 Hz (Fig. 7)
2. pulse width: 70 µs up to 500 µs (Fig. 8)
3. Ramp up: 1 to 30 s
4. Ramp down: 1 to 30 s
5. Work time: 1 to 30 s
6. Total run time: 1 up to 99 min. (Fig. 12)
Programme 21:
muscle stimulation
1. frequency: 20 Hz up to 120 Hz (Fig. 7)
2. pulse width: 60 µs up to 500 µs (Fig. 8)
3. Ramp up: 0,5 to 4 s
4. Ramp down: 0,5 to 2 s
5. Work time: 1 to 25 s
6. Total run time: 1 up to 99 min. (Fig. 12)
To save the new programme, press the
key once more after processing
the last parameter. An “U“ symbol now appears before the programme number
on the display.