Operating the SPORÉCUP XTR2 

 

Connecting cables and electrodes 

ƒ

 

Connect the electrodes with the cable(s) (always 2 electrodes per cable). 

ƒ

 

Connect the cable(s) to the XTR2 (no. 6). 

ƒ

 

Place the electrodes on the desired area (see examples on chapter Electrode Placement). 

 

Switching-ON the XTR2 

ƒ

 

Switch the commutator (no. 8) to the ON position. 

ƒ

 

Switch on the stimulator by pressing the  

z

  key (no. 5) 

The program shown on the display is the last which was used during the previous session.  

 

If the display shows a key symbol on the upper left side, the XTR2 has previously locked a 
specific program (see chapter locking the key pad). 
 

Selecting a program 

Select a program by pressing the 

P

 key. To scroll through the programs, press the 

P

 key as often 

as necessary until you have selected the program of your choice. To return to the previous 
program, press one of the 

 keys. Selecting a program is only possible if the keypad has not 

been interlocked on one specific program (see chapter interlocking the key pad). 
 

 

 

Setting the intensity 

Start the stimulation by increasing the intensity with the 

 key of the channel(s) connected to the 

electrodes.  
 

 key to increase the intensity 

 key to decrease the intensity 

 
The intensity of each channel is displayed on the screen. The intensity can be set between 0 and 
100 mA. By keeping the intensity keys pressed, the intensity will increase/decrease more rapidly. 

 

 

 

Terminating the stimulation 

You can terminate the stimulation at any time by pressing the 

P

 key. When the therapy time is 

completed the stimulation terminates automatically. If you press the  

z

  key the stimulation stops 

and the XTR2 switch off. 
 

Locking the key pad 

Choose the desired program by pressing the 

key (see chapter choosing a program). 

Press and hold for 3 seconds the 

P

 key and the right 

 keys to lock the stimulator on the chosen 

program. The XTR2 can now only be used with the chosen program. A key symbol appears on 
the display adjacent to the program number. All keys except the intensity and the ON/OFF keys 
are locked into this mode. Use the same key combination to unlock the stimulator.  
 

SPORÉCUP XTR2   |   Instruction Manual 

 

Содержание 104064

Страница 1: ...SPORÉCUP XTR2 MUSCLE STIMULATOR FOR SPORTS AND FITNESS Instruction Manual Art no 101478 ...

Страница 2: ...r 7 Switching OFF the stimulator 7 Charging the battery 7 Description of the programs 8 7 Setting of intensity levels 10 8 General information 10 9 Delivery Content 11 10 Accessories 11 Stimex self adhesive electrodes 12 Stimex FLOWER POWER self adhesive electrodes 12 Silicone electrodes to be used with contact gel 12 Contact gel for silicone electrodes item no 108000 Tube 60 g Fehler Textmarke ni...

Страница 3: ...erating close for example 1m to short wave and microwave units or mobile telephone systems can cause fluctuations in the output values of the electric current stimulation unit Do not use the machine during the simultaneous operation of other machines when driving or when asleep Contra indication Please check with your physician before using the XTR2 if one of these applies to you Patients with ele...

Страница 4: ... instructions manual 0197 Conformation with Council Directive 93 42 EEC of 14 June 1993 concerning medical devices BF type application part Protection against electric shocks SN The serial number of the product follows this symbol REF The article or order number of the product follows this symbol This equipment is marked with the recycling symbol It means that at the end of the working life of the...

Страница 5: ...ut short circuit control element AKS 30 integrated pre set programs Output current 100mA with an actual resistor of 1kΩ Frequency range 0 5 120 Hz Impulse width 50 400 μs Nominal current 15 mA Power supply integrated rechargeable accumulators Ni mH 4 8 V Dimensions 11 4 x 5 9 x 2 7 cm Weight approx 300 g 5 SPORÉCUP XTR2 Instruction Manual ...

Страница 6: ...chapter interlocking the key pad Setting the intensity Start the stimulation by increasing the intensity with the key of the channel s connected to the electrodes key to increase the intensity key to decrease the intensity The intensity of each channel is displayed on the screen The intensity can be set between 0 and 100 mA By keeping the intensity keys pressed the intensity will increase decrease...

Страница 7: ...omatically Charging the battery The battery level is shown on the display with a battery symbol with 4 lines If the battery level is too low the stimulator switches off automatically and cannot be turned on again The battery needs to be re charged Switch the commutator to the OFF position no 8 Connect the charger with the XTR2 socket no 7 Connect the charger to the mains circuit The diode on the c...

Страница 8: ...sms P2 Recovery after competition This program uses even a lower frequency than P1 in order to distress the tired muscles The blood flow is accelerated while the activity automatically slows down and the muscle relaxation is intensified The muscles recover better and faster for optimized future performance P3 Recovery after force training This program is recommended after an intensive force traini...

Страница 9: ...atica P14 Tendinitis P15 Joint pain P16 Chronic pain endorphinic SPORT P17 Resistance P18 Endurance P19 Strength P20 Explosive strength P21 Hypertrophy body building REHABILITATION P22 Amyotrophy prevention CAPILLARIZATION P23 Capillarization 1st phase P24 Capillarization 2nd phase FITNESS P25 Muscle starter P26 Toning muscle building ...

Страница 10: ...mall muscles Muscle fatigue A tired muscle will only tolerate a lower intensity Age of electrodes Self adhesive electrodes do not last forever You can use them up for up to 60 stimulation sessions This depends on the quality of the electrodes but also on the skin of the user Old electrodes have a higher resistance and their conductivity levels decline It is important to replace electrodes regularl...

Страница 11: ...n the XTR2 every 24 months This includes 1 Confirming that user instructions have been included in the accompanying documentation 2 Checking the equipment for completeness 3 Visual check for mechanical damage for damage to all cables and sockets 4 Functional Safety Checking the output signals with a load resistance of 1 kΩ real current and voltage Checking the frequency Checking the pulse width Th...

Страница 12: ... 283600 Stimex 50x90 mm 2 281007 Stimex 50x90 mm 4 283000 Stimex 50x130 mm 2 283100 Stimex 80x130 mm 2 281027 Stimex sensitive 50x50 mm 4 Stimex FLOWER POWER self adhesive electrodes Art No Description Quantity 281042 Flower Power 50 mm round 4 281035 Flower Power 50x50 mm 4 281040 Flower Power 50x90 mm 4 Silicone electrodes to be used with contact gel Art No Size Quantity 107090 20 mm round 2 107...

Страница 13: ...age with pelotte and integrated electrodes to be used with the XTR2 MedicoBack P type offers patients lumbar vertebrae stabilisation and support as well as easy nerve stimulation thanks to the integrated electrodes The pelotte with its massage nubs intensifies the pain relieving effect MedicoBack P type may be used for light or acute pain as well as for chronic pain in the lower back area Art No A...

Страница 14: ...12 Electrode placement Electrode placement for pain treatment Back pain Shoulder pain Epicondylitis ulnaris Epicondylitis radialis 14 SPORÉCUP XTR2 Instruction Manual ...

Страница 15: ...Arthrosis hip Kaada one canne only Sciatic pain Gonarthrosis Knee pain Ankle joint pain Trigeminus neuralgia 15 SPORÉCUP XTR2 Instruction Manual ...

Страница 16: ...16 SPORÉCUP XTR2 Instruction Manual Migraine Tension headache Polyarthritis Polyneuropahtia Carpal tunnel syndrome Pain of achilles tendon heel pain ...

Страница 17: ...requency parameters for an improved pain control or with low frequency parameters for optimized muscle stimulation This treatment method also has an enhanced effect on lymphatic drainage stimulation The placement of electrodes for dynamic stimulation depends on the direction of stimulation rising falling stimulation or sideways stimulation Rising or falling stimulation Sideways stimulation for bac...

Страница 18: ...Electrode placement for muscle stimulation Hand flebo Hand extensor Biceps Back Quadriceps Leg flebo hamstrings Foot flexor M tibialis Calf 18 SPORÉCUP XTR2 Instruction Manual ...

Страница 19: ...y building P6 Next day regeneration P7 Venous return dynamic stimulation REHABILITATION P8 Heavy legs P22 Amyotrophy prevention RELAXATION CAPILLARIZATION P10 Slow dynamic massage P23 Capillarization 1st phase P24 Capillarization 2nd phase PAIN TREATMENT FITNESS P11 Cervical pain P25 Muscle starter P12 Low back pain P26 Toning muscle building P13 Lumbosciatica P14 Tendinitis P15 Joint pain P16 Chr...

Страница 20: ...r GmbH Hoersnheimer Eck 19 35578 Wetzlar Germany www pierenkemper eu Distributer schwa medico GmbH Wetzlarerstr 41 43 35630 Ehringshausen Germany www schwa medico com 20 SPORÉCUP XTR2 Instruction Manual export schwa medico de ...

Отзывы: