J1034-UV-A/B/E DETECTOR Manual 

 

GENERAL DESCRIPTION 

 
This manual outlines the UV-A, UV-B, UV-E, UV-AB, UV-A/UV-E and UV-A/UV-B 
Broadband  Detectors  for  measuring  precise  global  UV  irradiance  for  different 
spectral  ranges.  The  system  provides  a  considerably  improved  measurement 
accuracy  in  comparison  to  the  traditional  detectors.  The  direct  cosine  error  f

2

  is 

typically less than 1.5%. 
The  detectors  are  flexible  enough  to  be  easily  installed  on  every  measurement 
site. Detector base plates for individual alignment work and a shadow ring system 
‘J1033-SCHADOWRING’ make an effective meteorological system. 

Technical Data 

 

UV-E 

UV-B 

UV-A 

UV-AB 

UV-A / 
UV-E 

UV-A / 
UV-B 

 

Relative spectral 
response 

ISO 

17166:1999/ 

CIE S 

007/E: 1998 

 

265-315 

nm 

 

310-400 

nm 

 

280-400 

nm

 

 

See individual 

detectors

 

Type 

Single band 

Dual band 

Cosine error f

2

 

<1.5% 

Cosine Response 

 

< 2.5% between 0°and 70° solar zenith angle

 

Measure range 

0 – 0.6 

W/m² 

0 - 5 

W/m² 

0 - 100 

W/m² 

0 – 100 

W/m² 

0 – 100  

0 – 0.5 

W/m² 

0 – 100  

0 – 0.5 

W/m² 

Signal output 

0 - 5 V  (on request: 0 - 2.5 V) 

Temperature range 

-40 °C ÷ +60 °C 

Temp. coefficient 

0.1 % / K 

Power Supply 

9 - 24 V / 0.75mA 

Time switch on 

< 1 sec 

Linearity 

< 1% 

Housing 

Anodised aluminium, Quartz glass dome 

Dimension 

80 mm (diameter D) x 88mm (height H) 

Leveling plate: 115 mm (D) x 88mm (H) 

Weight 

< 1 kg 

 

Pin connection 

 

1: Vcc (white) 
2: GND (brown) 
3: UV-E

AE

 

4: UV-A

AE

 

5: Housing (screen) 

Pin connection of all UV detectors 

1 – Vcc 

[white] 

9 - 24 V / 0.75mA 

2 – GND 

[brown] 

Ground for power supply and signal output 

3 – N.C./Output UV  [gray] 

not connected (single band) 
UV-E (dual band) output 

4 – Output UV 

[yellow] 

UV-B, UV-A, UV-E, UV-AB output 
UV-A (dual band) output 

5 – Housing  

  

 

Содержание J1034-UV-A

Страница 1: ...34 UV A B E DETECTOR USER S MANUAL C M S In g D r Sc h re de r G m b H Lofererstraße 32 A 6322 Kirchbichl Tel 43 0 5332 77056 00 Fax 43 0 5332 77056 14 E Mail info schreder cms com Web www schreder cms com ...

Страница 2: ...ake one copy of the software disk for backup purposes The software may not be copied or distributed in any other way No part of this manual may be reproduced or transmitted in any form or by any means electronic optical or mechanical including photocopying and recording or by any information storage and retrieval system without permission in writing from CMS Ing Dr Schreder GmbH ...

Страница 3: ...1 WARRANTY 1 CONTACT INFORMATION 1 SAFETY INFORMATION 2 WARNING CAUTION 2 VOLTAGE WARNING 2 GENERAL DESCRIPTION 3 TECHNICAL DATA 3 PIN CONNECTION OF ALL UV DETECTORS 3 RELATIVE SPECTRAL RESPONSE 4 MOUNTING AND ALIGNMENT 5 COMPONENTS PARTS REFERENCE 6 LIST OF ADDITIONAL PARTS 6 REFERENCES 6 ...

Страница 4: ...s the appropriate specifications The product warranty is valid only if the product has been installed and used according to the directives provided in this instruction manual CMS shall in no event be liable for incidental or consequential damages arising from the faulty and incorrect use of the product In case of any manufacturing defect the product will be repaired or replaced under warranty Howe...

Страница 5: ...ation base should have enough load capacity for the instrument to be mounted Fix the equipment securely Otherwise the instrument may drop due to gale or earthquake which may lead to unexpected accidents Make sure the instrument and the cables are installed in a location where they will not get soaked Voltage Warning Power Supply Make sure to check the power supply voltage and type AC DC before con...

Страница 6: ...6 1999 CIE S 007 E 1998 265 315 nm 310 400 nm 280 400 nm See individual detectors Type Single band Dual band Cosine error f2 1 5 Cosine Response 2 5 between 0 and 70 solar zenith angle Measure range 0 0 6 W m 0 5 W m 0 100 W m 0 100 W m 0 100 0 0 5 W m 0 100 0 0 5 W m Signal output 0 5 V on request 0 2 5 V Temperature range 40 C 60 C Temp coefficient 0 1 K Power Supply 9 24 V 0 75mA Time switch on...

Страница 7: ...ure 1 Relative Spectral Response of the single band detectors Erythemal weighting function ERY defined ISO 17166 1999 CIE S 007 E 1998 260 280 300 320 340 360 380 400 10 4 10 3 10 2 10 1 10 0 WAVELENGTH nm RELATIVE SPECTRAL RESPONSE ERY UV E AE UV A AE Figure 2 Relative Spectral Response of the dual band detector Erythemal weighting function ERY defined ISO 17166 1999 CIE S 007 E 1998 ...

Страница 8: ... to the UV detector The additional available Detectorbase 105303 is a general purpose base plate which fits directly to the J1033 Shadowring 103304 It allows a robust and quick alignment to individual grounds Use the tilting screws to align the detector and control the correct settings with the help of the spirit level After that use the montage screws to fix the detector to the ground Figure 3 UV...

Страница 9: ...AW Report No 141 Report of the LAP COST WMO Intercomparison of Erythemal Radiometers Thessaloniki Greece 1999 WMO TD No 1051 2 WMO GAW Report No 146 Quality Assurance in Monitoring Solar Ultraviolet Radiation the State of the Art WMO TD No 1180 3 WMO GAW Report No 164 Instruments to measure Ultraviolet Radiation Part 2 Broadband Instruments Measuring Erythemally Weighted Solar Radiation WMO TD No ...

Отзывы: