Instructions for Use SS 20.200 

Page 4 

3  Mounting instructions 

The 

SS  20.200

  is  a  high-precision  and  sensitive  measuring  device.  Ac-

cordingly, the mechanical stress on the probe head must be minimised in 
all mounting steps. 

 

Avoid mechanical stress of the sensor tip as much as possible, 
since otherwise irreversible damage can occur. 

Mounting position 

In  order  to  achieve  maximum  measuring  sensitivity,  the  sensor  head 
should  be  mounted  pointing  upwards.  A  sensor  head  pointing  down-
wards must be avoided, since this will increase the measuring range limit 
(nominal value 0.06 m/s). 

In order to obtain good measurement results, the flow measuring sleeve 
must always be in the centre of the flow duct. 

 

The sensor head must immerse completely (that is, including 
the temperature measuring sleeve) into the air flow. If this is not 
the case, the measured value can be completely wrong. 

The  sensor  should  not  be  mounted  in  the  proximity  of  heat  sources, 
since  this  may  result  in  measurement  errors.  When  used  in  ventilation 
pipes  equipped  with  heat  exchangers,  the  sensor  should  always  be 
mounted upstream of the heat exchanger. 

Wall mounting 

First  the  protective  cover  must  be  pulled  off  the  sensor  head.  Then  the 
sensor  together  with  the  threaded  piece  (M  18  x  1  fine  pitch  thread)  is 
screwed  in  the  appropriate  threaded  bore  of  the  duct  wall.  Set  the  de-
sired position and secure it with a counter nut. 

Alternatively 

SCHMIDT  Technology

  offers  auxiliary  installation  equip-

ment. Please refer to following URLs for details: 

www.schmidt-sensors.com

 

or 

www.schmidttechnology.com

  

This  also  includes  through-bolt  joints  that  are  suitable  for  use  in  media 
with overpressure

2

 in which the following has to be considered: 

 

Mounting and dismounting of the sensor can be carried out only 
as long as the system is in 

depressurized state

Only using proper pressure-tight seals (e. g. Teflon tape). 
Appropriate safety precautions are taken to avoid unintended 
discarding of the sensor due to overpressure. 

                                            

2

 Pressure-tight, including pressure security kit (material no. 524891 and 524919) 

!

!

!

Содержание SS 20.200

Страница 1: ...SCHMIDT Flow Switch SS 20 200 Instructions for Use ...

Страница 2: ...mation 3 2 Application range 3 3 Mounting instructions 4 4 Electrical connection 6 5 Commissioning 7 6 Switching threshold 8 7 Service information 9 8 Technical data 10 9 Certificate of conformity 11 Imprint Copyright 2016 SCHMIDT Technology All rights reserved Version 505518 01C Subject to modifications ...

Страница 3: ... the delivery performance or use of this unit Symbols used in this manual The symbols used in this manual are explained in the following section Danger warnings and safety instructions read them care fully Non observance of these instructions may lead to injury of the personnel or malfunction of the device 2 Application range The SCHMIDT Flow Switch SS 20 200 has been designed for station ary use ...

Страница 4: ... heat sources since this may result in measurement errors When used in ventilation pipes equipped with heat exchangers the sensor should always be mounted upstream of the heat exchanger Wall mounting First the protective cover must be pulled off the sensor head Then the sensor together with the threaded piece M 18 x 1 fine pitch thread is screwed in the appropriate threaded bore of the duct wall S...

Страница 5: ...b ances opposite to the flow direction L1 L2 L D L Total length of measuring distance L2 Length of run out distance L1 Length of run in distance D Diameter of the measuring distance The following table shows the necessary straight conduit lengths as a function of the pipe diameter D for various scenarios Flow obstacle upstream of measuring conduit Minimum length of run in distance L1 Minimum lengt...

Страница 6: ...ble assignment The sensor is equipped with a cable firmly fixed to its housing with 2 m length and open cable ends pin assignments see following graphic Sensor 24V R1 RON R2 Gnd WH white YE yellow GN green Z During electrical installation ensure that no voltage is applied and inadvertent activation is not possible For signalling a semiconductor relay is used Its galvanically decoupled switching ou...

Страница 7: ...Switching threshold1 underrun exceeded Closed Open Sensor ready Switching threshold3 exceeded underrun Open Closed Legend LED off LED flashing 2 Hz LED on Startup Before switching on it must be checked if the SS 20 200 is installed cor rectly both mechanically and electrically The sensor is ready within 20 s after applying the operating voltage During this initialisation stage the green LED is fla...

Страница 8: ...nge The individual setting of the switching point is done by adjusting the flow velocity to the desired threshold value Then the potentiometer is rotated from the zero point clockwise until the relay is switching The release point is then offset downward by the hysteresis see also chapter Hyste resis of switching threshold For fine adjustment the potentiometer can now be rotated back a little bit ...

Страница 9: ...sor head is a sensitive measuring system During manual cleaning proceed with great care Before putting it again into operation wait until the sensor head is com pletely dry Sterilization Both uncoated and coated sensors can be sterilized during operation Alcohols drying without leaving residues and hydrogen peroxide un coated version only are approved and certified disinfectants Spare parts or rep...

Страница 10: ...r probe 20 85 C Electronics 20 70 C Storage temperature 20 85 C Humidity range 0 95 of rel humidity RH none condensing Operating pressure 0 10 bar overpressure Operating voltage UB 24 VDC 20 Current consumption typ 40 mA max 70 mA 5 Switching relay 30 VPeak 100 mA 300 mW Display LED Green sensor status Red switching status sensor defective Electrical connection Non detachable cable pigtail6 4 x 0 ...

Страница 11: ...Instructions for Use SS 20 200 Page 11 9 Certificate of conformity ...

Страница 12: ...uctions for Use SS 20 200 Page 12 SCHMIDT Technology GmbH Feldbergstraße 1 78112 St Georgen Germany Phone 49 0 7724 899 0 Fax 49 0 7724 899 101 Email sensors schmidttechnology de URL www schmidt sensors com ...

Отзывы: